Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient...
Philips North America began a recall of Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system on . FDA classified it as Class II and gives the reason as "Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0975-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Beginning in January 2022, Philips Field Service Engineers (FSE) visited impacted customer sites to replace the magnetic castor wheels of the HA Flextrack Trolley II with non-magnetic castor wheels. All reachable systems were addressed as of August 10, 2023.
- Quantity in commerce
- 10
- Distribution
- Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
- Product codes and lots
- UDI-DI 00884838090040 Ingenia Serial number / Accessory Serial Number 47004 100020 47019 100024 47190 100022 47200 100025 47210 100019 47227 100063 47231 100056 47271 100032 47281 100028 47281 100031
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0973-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0974-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0975-2024 | Class II | Ongoing | Voluntary: Firm initiated |