Z-0835-2024Class II

Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion...

Philips Medical Systems Nederland B.V. / initiated 2023-12-12 / Open, Classified

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion 3M12 -722063, 722221; Azurion 3M15 - 722064... on . FDA classified it as Class II and gives the reason as "System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0835-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips Medical Systems Nederland B.V.
Product
Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion 3M12 -722063, 722221; Azurion 3M15 - 722064, 722222; Azurion 5M12 - 722227; Azurion 5M20 - 722228; Azurion 7B12/12 - 722067, 722225; Azurion 7B20/15 - 722068, 722226; Azurion 7M12 - 722078, 722223; Azurion 7M20 - 722079, 722224; Azurion 3 M15 (China) - 722280;
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.
FDA root cause
Under Investigation by firm
Action
On December 12, 2023, the firm notified customers via certified mailing of URGENT Medical Device Correction letters. Customers may identify affected products by the system product name and model number, which can be found on the System Identification Label located on the system stand. The software version of the Philips Azurion system can be identified during start-up. Customers were advised to keep the number of studies in the patient database as low as possible, but at least below 500. Customers should keep the Urgent Medical Device Correction letter with the documentation of the system until Philips corrects the system. Ensure that all users of the system are notified of the issue. Philips is working on a software release that will correct this issue. Philips representatives will contact affected customers to schedule the correction. Please contact your local Philips representative for further information or support concerning this issue.
Quantity in commerce
1,799 US; 5,747 OUS
Distribution
Worldwide - US Nationwide distribution.
Product codes and lots
System Product Name - UDI-DI Azurion 3M12 - 00884838085275, 00884838099203, 00884838101890; Azurion 3M15 - 00884838085282, 00884838099210; Azurion 5M12 - 00884838099227; Azurion 5M20 - 00884838099234; Azurion 7B12/12 - 00884838059290, 00884838064775, 00884838085350, 00884838099265; Azurion 7B20/15 - 00884838059207, 00884838064775, 00884838085367, 00884838099272, 00884838101883; Azurion 7M12 - 00884838085251, 00884838099241, 00884838101890; Azurion 7M20 - 00884838059184, 00884838059191, 00884838059207, 00884838059221, 00884838059290, 00884838064775, 00884838085268, 00884838086777, 00884838099258, 00884838101883, 00884838101890; Azurion 3 M15 (China) - 00884838085282, 00884838099210, 00884838103276; Software Version Numbers R1.0, R1.1, R1.2, R2.0, R2.1, and R2.2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0835-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0835-2024Class IIOngoingVoluntary: Firm initiated