BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791...

Philips North America LLC / initiated 2023-12-15 / Open, Classified / root cause: design

Philips North America began a recall of BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 453560279801; 4x4 digit format... on . FDA classified it as Class I and gives the reason as "Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0929-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 453560279801; 4x4 digit format 2170-3000A 2170-3001A 2170-3002A 2170-3003A
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
Root cause tag
design
FDA root cause
Component design/selection
Action
On December 15, 2023, the firm notified affected customers via URGENT Medical Device Correction letters. Customers were instructed to not position a patient's lower limbs directly under the detector below the Center of Gantry bore. Customers should circulate the notice to all users of the device for awareness and retain the letter with affected systems until the correction is complete. Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit the customer site and correct the system if necessary. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377). As of 3/14/24, Philips will begin inspecting all affected devices in the field to ensure alignment of the radial leadscrew. Systems that are identified as having a misaligned leadscrew will undergo a report. Additionally, Philips will begin intsallation of a secondary catching mechanism (safety plate) starting in May 2024 in all devices. The firm will contact customers to coordinate these activities.
Quantity in commerce
302 US; 264 OUS
Distribution
Domestic distribution nationwide. International distribution worldwide.
Product codes and lots
All serial numbers are affected. No UDI-DI available for this product.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0929-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0929-2024Class IOngoingVoluntary: Firm initiated
Z-0930-2024Class IOngoingVoluntary: Firm initiated
Z-0931-2024Class IOngoingVoluntary: Firm initiated