BrightView X, gamma camera for SPECT; Model Nos.: 6-digit format 882478; 12-digit format 453560824741...
Philips North America began a recall of BrightView X, gamma camera for SPECT; Model Nos.: 6-digit format 882478; 12-digit format 453560824741 453560829261 on . FDA classified it as Class I and gives the reason as "Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0930-2024
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- BrightView X, gamma camera for SPECT; Model Nos.: 6-digit format 882478; 12-digit format 453560824741 453560829261;
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On December 15, 2023, the firm notified affected customers via URGENT Medical Device Correction letters. Customers were instructed to not position a patient's lower limbs directly under the detector below the Center of Gantry bore. Customers should circulate the notice to all users of the device for awareness and retain the letter with affected systems until the correction is complete. Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit the customer site and correct the system if necessary. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377). As of 3/14/24, Philips will begin inspecting all affected devices in the field to ensure alignment of the radial leadscrew. Systems that are identified as having a misaligned leadscrew will undergo a report. Additionally, Philips will begin intsallation of a secondary catching mechanism (safety plate) starting in May 2024 in all devices. The firm will contact customers to coordinate these activities.
- Quantity in commerce
- 56 US; 46 OUS
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
- Product codes and lots
- All serial numbers are affected. No UDI-DI available for this product.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0929-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0930-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0931-2024 | Class I | Ongoing | Voluntary: Firm initiated |