Achieva 3.0T. Model. (REF) Numbers 781277, 781177, 781278, 781344, 781345
Philips North America began a recall of Achieva 3.0T. Model. (REF) Numbers 781277, 781177, 781278, 781344, 781345 on . FDA classified it as Class II and gives the reason as "The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0948-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- Achieva 3.0T. Model. (REF) Numbers 781277, 781177, 781278, 781344, 781345.
- Product code
- LNI System, Nuclear Magnetic Resonance Spectroscopic
- Reason for recall
- The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On January 4, 2024, the firm began notification of affected customers via URGENT Medical Device Correction letters. Customers were provided with precautions to take when using the device, including inspection of the QBC seal for separation. If the QBC seal is found to be loose, or becomes loose during a patient scan, customers should immediately stop use and contact their local Philips service representative. Customers should also ensure that all users of the device are aware of the issue and retain the letter with affected systems until a solution is installed. Philips will contact customers to schedule a time for a Field Service Engineer (FSE) to visit your site and replace your system's QBC Seal. Philips plans to start implementing corrections in Q3 2024. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377). Revision B of the URGENT Medical Device Correction letter was distributed to the additional US consignees on 24-Mar-2025. It was later identified that UMDC Rev B contained three MR systems with a 70cm bore (Ingenia 1.5T: 781341, Ingenia 3.0T: 781377, Ingenia Ambition X: 782109) which were incorrectly included, as the scope of this field action is MR systems with 60cm bore. The URGENT Medical Device Correction letter was updated to Rev C to remove these systems. Revision C of the UMDC was distributed to the additional US consignees on 09-Apr-2025.
- Quantity in commerce
- 75 US; 442 OUS
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- a. Model No. 781277; UDI-DI N/A. b. Model No. 781177; UDI-DI N/A. c. Model No. 781278; UDI-DI 00884838004139. d. Model No. 781344; UDI-DI N/A. e. Model No. 781345; UDI-DI N/A. All units affected
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0945-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0946-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0947-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0948-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0949-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0950-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0951-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0952-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0953-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0954-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0955-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0956-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.