SmartPath to dStream for 1.5T, Model No. 781260, 782112
Philips North America began a recall of SmartPath to dStream for 1.5T, Model No. 781260, 782112 on . FDA classified it as Class III and gives the reason as "An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0812-2024
- Classification
- Class III
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- SmartPath to dStream for 1.5T, Model No. 781260, 782112
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- On December 29, 2023, the firm notified affected customers with URGENT Medical Device Correction letters. The firm stated there is no impact to patients or users. There is a potential risk of electrical shock for service personnel. Customers may continue to use MR systems in accordance with the intended use. Circulate this notice to all users of this device so that they are aware of the issue, and post the letter in the technical room until the electrical shock warning label is updated. A Philips representative will contact you to schedule a time for a Field Service Engineer to verify the electrical shock warning label and apply the correct label if applicable. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
- Quantity in commerce
- 7 US; 16 ROW total
- Distribution
- Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.
- Product codes and lots
- Model No. 781260 UDI-DI 00884838095076 Serial Numbers 21020 9024 8657 21200 21223 18948 32131; Model No. 782112 UDI-DI 00884838098886 Serial Numbers 30089 20204
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0807-2024 | Class III | Ongoing | Voluntary: Firm initiated | ||
| Z-0808-2024 | Class III | Ongoing | Voluntary: Firm initiated | ||
| Z-0809-2024 | Class III | Ongoing | Voluntary: Firm initiated | ||
| Z-0810-2024 | Class III | Ongoing | Voluntary: Firm initiated | ||
| Z-0811-2024 | Class III | Ongoing | Voluntary: Firm initiated | ||
| Z-0812-2024 | Class III | Ongoing | Voluntary: Firm initiated |