Z-0812-2024Class III

SmartPath to dStream for 1.5T, Model No. 781260, 782112

Philips North America LLC / initiated 2023-12-29 / Open, Classified / root cause: labeling and UDI

Philips North America began a recall of SmartPath to dStream for 1.5T, Model No. 781260, 782112 on . FDA classified it as Class III and gives the reason as "An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0812-2024
Classification
Class III
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
SmartPath to dStream for 1.5T, Model No. 781260, 782112
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
On December 29, 2023, the firm notified affected customers with URGENT Medical Device Correction letters. The firm stated there is no impact to patients or users. There is a potential risk of electrical shock for service personnel. Customers may continue to use MR systems in accordance with the intended use. Circulate this notice to all users of this device so that they are aware of the issue, and post the letter in the technical room until the electrical shock warning label is updated. A Philips representative will contact you to schedule a time for a Field Service Engineer to verify the electrical shock warning label and apply the correct label if applicable. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Quantity in commerce
7 US; 16 ROW total
Distribution
Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.
Product codes and lots
Model No. 781260 UDI-DI 00884838095076 Serial Numbers 21020 9024 8657 21200 21223 18948 32131; Model No. 782112 UDI-DI 00884838098886 Serial Numbers 30089 20204

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0812-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0807-2024Class IIIOngoingVoluntary: Firm initiated
Z-0808-2024Class IIIOngoingVoluntary: Firm initiated
Z-0809-2024Class IIIOngoingVoluntary: Firm initiated
Z-0810-2024Class IIIOngoingVoluntary: Firm initiated
Z-0811-2024Class IIIOngoingVoluntary: Firm initiated
Z-0812-2024Class IIIOngoingVoluntary: Firm initiated