Z-1037-2024Class II

BD BBL Sensi Disc Piperacillin/Tazobactam- 100/10 ug, used for semi-quantitative in vitro susceptibility...

Becton Dickinson & Co / initiated 2024-01-08 / Open, Classified / root cause: manufacturing process

Becton Dickinson & Co. began a recall of BD BBL Sensi Disc Piperacillin/Tazobactam- 100/10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231691 and 231692 on . FDA classified it as Class II and gives the reason as "There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1037-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Becton Dickinson & Co
Product
BD BBL Sensi Disc Piperacillin/Tazobactam- 100/10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231691 and 231692
Product code
JTN Susceptibility Test Discs, Antimicrobial
Reason for recall
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On January 8, 2024, the firm notified customers through URGENT Medical Device Product Correction letters. Users were instructed to cease use of catalog number 291270, and to refrain from conducting AST for H. influenzae with impacted Sensi-Discs. Customers should destroy any units of catalog number 291270 in their inventory. Customers should share and post the recall notification within their facility network and forward to any customers to whom they may have distributed the product. Future lots of affected Sensi-Disc products (excluding catalog number 291270) will be labeled with a sticker notifying customers "This product should not be used for the semi-quantitative in vitro susceptibility testing of Haemophilus influenzae." Identify impacted lot numbers: https://legacy.bd.com/alerts-notices/IDS-23-4851.asp For customer questions: Phone: 1-844-823-5433 Say "Recall when prompted Mon-Fri 8:00am and 5:00pm CT or Email: [email protected]
Quantity in commerce
2,363,168 total units
Distribution
Domestic distribution nationwide. International distribution worldwide.
Product codes and lots
Catalog No. 231691 UDI-DI 00382902316916 Lots 1301719 2026369 2056587 2119595 3003951 3062107 3118199 3275730; Catalog No. 231692¿ UDI-DI 30382902316924 Lots 3003951 3184117 3223408 3275730 1301719 2026369 2056587 2119595 2147716 2244463 3062107 3118199

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1037-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1011-2024Class IIOngoingVoluntary: Firm initiated
Z-1012-2024Class IIOngoingVoluntary: Firm initiated
Z-1013-2024Class IIOngoingVoluntary: Firm initiated
Z-1014-2024Class IIOngoingVoluntary: Firm initiated
Z-1015-2024Class IIOngoingVoluntary: Firm initiated
Z-1016-2024Class IIOngoingVoluntary: Firm initiated
Z-1017-2024Class IIOngoingVoluntary: Firm initiated
Z-1018-2024Class IIOngoingVoluntary: Firm initiated
Z-1019-2024Class IIOngoingVoluntary: Firm initiated
Z-1020-2024Class IIOngoingVoluntary: Firm initiated
Z-1021-2024Class IIOngoingVoluntary: Firm initiated
Z-1022-2024Class IIOngoingVoluntary: Firm initiated
Z-1023-2024Class IIOngoingVoluntary: Firm initiated
Z-1024-2024Class IIOngoingVoluntary: Firm initiated
Z-1025-2024Class IIOngoingVoluntary: Firm initiated
Z-1026-2024Class IIOngoingVoluntary: Firm initiated
Z-1027-2024Class IIOngoingVoluntary: Firm initiated
Z-1028-2024Class IIOngoingVoluntary: Firm initiated
Z-1029-2024Class IIOngoingVoluntary: Firm initiated
Z-1030-2024Class IIOngoingVoluntary: Firm initiated
Z-1031-2024Class IIOngoingVoluntary: Firm initiated
Z-1032-2024Class IIOngoingVoluntary: Firm initiated
Z-1033-2024Class IIOngoingVoluntary: Firm initiated
Z-1034-2024Class IIOngoingVoluntary: Firm initiated
Z-1035-2024Class IIOngoingVoluntary: Firm initiated
Z-1036-2024Class IIOngoingVoluntary: Firm initiated
Z-1037-2024Class IIOngoingVoluntary: Firm initiated
Z-1038-2024Class IIOngoingVoluntary: Firm initiated
Z-1039-2024Class IIOngoingVoluntary: Firm initiated
Z-1040-2024Class IIOngoingVoluntary: Firm initiated