Z-1087-2024Class II

CombiDiagnost R90 1.0 (709030)

Philips North America LLC / initiated 2022-02-11 / Open, Classified / root cause: software

PHILIPS MEDICAL SYSTEMS began a recall of CombiDiagnost R90 1.0 (709030) on . FDA classified it as Class II and gives the reason as "While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1087-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
CombiDiagnost R90 1.0 (709030)
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.
Root cause tag
software
FDA root cause
Software design
Action
An Important Product Notice (2021-PD-DXR-034) dated February 15, 2022 was issued via Certified Mail. Philips Field Service Engineers will contact customers to schedule an onsite visit to install update software correcting the issue. Customers are asked to follow the provided instructions to ensure there are no pending images before an exam circulate the notice to all users and to complete and return the response form and circulate the notice to all users of the device for awareness. Contact Philips Customer Care Solutions Center 1-800-722-9377 with questions.
Quantity in commerce
86 units
Distribution
US Nationwide distribution in the states of CA, FL, GA, IL, IN, KS, MD, ME, MI, MN, NC, NJ, NY, TX, VA, WA.
Product codes and lots
Model: R90 1.0 (709030); UDI DI: 00884838076747; Software version 1.1.1 or higher; Serial numbers: 10000000, 10000002, 10000003, 10000004, 10000006, 10000011, 10000012, 10000020, 10000024, 10000027, 10000049, 10000050, 10000065, 10000066, 10000067, 10000069, 10000078, 10000081, 10000099, 10000103, 10000107, 10000140, 10000145, 10000152, 10000180, 10001001, 10001010, 10001095, 591362, SN16000009, SN17000003, SN17000007, SN17000015, SN17000022, SN17000111, SN17000114, SN17000118, SN17000120, SN17000121, SN17000122, SN17000123, SN17000128, SN17000129, SN17000130, SN17000145, SN17000146, SN17000147, SN17000148, SN17000153, SN17000157, SN17000163, SN17000165, SN17000166, SN18000003, SN18000004, SN18000010, SN18000021, SN18000025, SN18000033, SN18000037, SN18000038, SN18000044, SN18000045, SN18000053, SN18000054, SN18000057, SN18000060, SN18000068, SN18000069, SN18000079, SN19000004, SN19000006, SN19000011, SN19000020, SN19000034, SN19000036, 27, 40, 41, 42, 77, 101, 131, 155, 180, 190;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1087-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1087-2024Class IIOngoingVoluntary: Firm initiated
Z-1088-2024Class IIOngoingVoluntary: Firm initiated