CombiDiagnost R90 1.1 (709031)
PHILIPS MEDICAL SYSTEMS began a recall of CombiDiagnost R90 1.1 (709031) on . FDA classified it as Class II and gives the reason as "While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1088-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- CombiDiagnost R90 1.1 (709031)
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- An Important Product Notice (2021-PD-DXR-034) dated February 15, 2022 was issued via Certified Mail. Philips Field Service Engineers will contact customers to schedule an onsite visit to install update software correcting the issue. Customers are asked to follow the provided instructions to ensure there are no pending images before an exam circulate the notice to all users and to complete and return the response form and circulate the notice to all users of the device for awareness. Contact Philips Customer Care Solutions Center 1-800-722-9377 with questions.
- Quantity in commerce
- 130 units
- Distribution
- US Nationwide distribution in the states of CA, FL, GA, IL, IN, KS, MD, ME, MI, MN, NC, NJ, NY, TX, VA, WA.
- Product codes and lots
- Model: R90 1.1 (709031); UDI DI: 00884838101456; Software version 1.1.1 or higher; Serial Numbers: 10001004, 10001005, 10001006, 10001008, 10001009, 10001010, 10001011, 10001012, 10001013, 10001018, 10001020, 10001022, 10001024, 10001026, 10001027, 10001028, 10001029, 10001030, 10001031, 10001032, 10001035, 10001036, 10001043, 10001044, 10001045, 10001046, 10001047, 10001048, 10001049, 10001050, 10001051, 10001052, 10001053, 10001054, 10001055, 10001056, 10001057, 10001058, 10001059, 10001060, 10001061, 10001062, 10001063, 10001064, 10001065, 10001066, 10001067, 10001068, 10001069, 10001070, 10001071, 10001072, 10001073, 10001074, 10001075, 10001076, 10001077, 10001078, 10001079, 10001080, 10001081, 10001083, 10001084, 10001085, 10001086, 10001087, 10001088, 10001089, 10001090, 10001091, 10001092, 10001093, 10001094, 10001095, 10001096, 10001097, 10001098, 10001099, 10001100, 10001101, 10001102, 10001103, 10001104, 10001105, 10001106, 10001107, 10001108, 10001109, 10001110, 10001111, 10001112, 10001113, 10001114, 10001115, 10001116, 10001117, 10001118, 10001119, 10001120, 10001121, 10001122, 10001123, 10001124, 10001125, 10001126, 10001127, 10001128, 10001129, 10001130, 10001131, 10001132, 10001133, 10001134, 10001135, 10001136, 10001137, 10001138, 10001139, 10001140, 10001141, 10001142, 10001143, 10001144, 10001145, 10001147, 10001148, 10001149, 10001150, 10001152, 10001201;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1087-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1088-2024 | Class II | Ongoing | Voluntary: Firm initiated |