3:1 Dermacarrier, Model Number 00219501300, skin graft carrier
Zimmer Surgical Inc began a recall of 3:1 Dermacarrier, Model Number 00219501300, skin graft carrier on . FDA classified it as Class II and gives the reason as "Due to manufacturing issue, the product is labeled as 3:1 ratio, but the ridge pattern is a 1.5:1 ratio. The problem may lead to minor tissue damage or additional unplanned graft if deemed necessary". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1078-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Zimmer Surgical SA
- Product
- 3:1 Dermacarrier, Model Number 00219501300, skin graft carrier
- Product code
- FZW Expander, Surgical, Skin Graft
- Reason for recall
- Due to manufacturing issue, the product is labeled as 3:1 ratio, but the ridge pattern is a 1.5:1 ratio. The problem may lead to minor tissue damage or additional unplanned graft if deemed necessary.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- On January 2, 2024, the firm notified affected customers of the issue through URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to immediately notify affected personnel of the recall. Customers should also immediately locate and quarantine affected product in inventory. The customer may return the product to Zimmer Biomet, or a Zimmer Biomet sales representative may remove and return the product on the customer's behalf. If product has been further distributed, please notify your customers of the recall and ensure documentation. If you have further questions or concerns after reviewing this notice, please call Zimmer Biomet's customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].
- Quantity in commerce
- 76 boxes (1,520 units) US; 521 boxes (10,420 units) OUS
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, DE, FL, GA, IA, IL, IN, KS, LA, MI, MO, MT, NC, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI, WV and the countries of AUTRALIA, CANADA, DNIPRO OBLAS, HONG KONG, INDIA, JAPAN, MEXICO, CDMX, NETHERLANDS, SEOUL KOREA, SINGAPORE, TAIWAN, THAILAND.
- Product codes and lots
- UDI-DI 00889024378780 Lots 65292843 (Expires 16-Oct-2026) and 65390419 (Expires 18-Jan-2027)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1078-2024 | Class II | Ongoing | Voluntary: Firm initiated |