Z-1009-2024Class II

Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope...

OLYMPUS Corporation of the Americas / initiated 2024-01-16 / Open, Classified / root cause: manufacturing process

Olympus Corporation of the Americas began a recall of Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent contamination of the rinse water... on . FDA classified it as Class II and gives the reason as "Water filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1009-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent contamination of the rinse water. The MAJ-2318 water filter is compatible with OER-ELITE and OER-MINI.
Product code
FEB Accessories, Cleaning, For Endoscope
Reason for recall
Water filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On January 16, 2024, the firm notified customers of the issue via URGENT MEDICAL DEVICE FIELD CORRECTIVE ACTION letters. Customers were instructed to immediately assess all MAJ-2318 at their facility to check the model and lot number. Customers were given detailed instructions on how to identify affected product, including for MAJ-2318 installed in the OER machine. Please contact your Olympus customer service representative at 1-800-848-9024, option 2 to obtain a Return Material Authorization to return any affected product. Customers who have further distributed product should notify downstream accounts at once. If you require assistance with identifying affected product, please contact the Olympus Technical Assistance Center at 1- 800-848-9024, option 1. If you have questions about this recall, please contact the recalling firm at (647) 999-3203 or by e-mail at [email protected].
Quantity in commerce
877 MAJ-2318, 119 OER-Elite, 3 OER-MINI
Distribution
Domestic US distribution nationwide.
Product codes and lots
UDI-DI 04953170401473 Lots 36A, 37A, 38A Affected product may have also been packaged within OER boxes (OER-Elite and OER-MINI)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1009-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1009-2024Class IIOngoingVoluntary: Firm initiated