BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078. Used with the CARDIOHELP...
Maquet Medical Systems USA began a recall of BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078. Used with the CARDIOHELP heart-lung machine on . FDA classified it as Class II and gives the reason as "The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1077-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- BEQ-HLS 7050 USA, HLS Set Advanced 7.0, Product Code/Part Number 70106.9078. Used with the CARDIOHELP heart-lung machine.
- Product code
- DTZ Oxygenator, Cardiopulmonary Bypass
- Reason for recall
- The Emergency Priming Line, a component of the HLS Set, may have a compromised sterile barrier due to nonconformities at the bag seal. This issue may result in inflammation, sepsis, or infection.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On January 16, 2024, the recalling firm began notifying affected customers via URGENT MEDICAL DEVICE CORRECTION letters. Customers were instructed to dispose of any Emergency Priming Line from a previously opened, affected HLS set. Customers should only use the Emergency Priming Line from a newly opened HLS Set. Device users should always have an HLS Set ready as a backup during therapy. The remaining components of the backup HLS Set must be disposed of after extraction of the Emergency Priming Line. Please reach out to Getinge Customer Support for the issuance of a replacement HLS Set at no cost. Customers may also return unopened/unexpired affected HLS Sets by contacting Customer Support. If an affected Emergency Priming Line is already in use, the firm advises it should remain in use due to increased potential risk of stopping this emergency procedure. Users should monitor the patient closely for signs of infection. Contact Getinge Customer Support: (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). ***Updated 3/21/24*** Additional lots identified. Firm notified affected customers on March 21, 2024.
- Quantity in commerce
- 16520 worldwide; 6657 US
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- UDI-DI 04058863080383 Lots 3000280906, 3000280907, 3000280908, 3000280909, 3000282253, 3000282254, 3000282255, 3000291127, 3000300688, 3000314506, 3000316206, 3000316600, 3000317531, 3000317532, 3000318850, 3000318851, 3000318852, 3000318853, 3000319748, 3000319749, 3000319750, 3000319751, 3000319778, 3000319781, 3000323082, 3000323084, 3000323085, 3000324495, 3000324496, 3000324497, 3000324499, 3000325557, 3000325558, 3000325559, 3000327120, 3000327121, 3000328404, 3000328405, 3000334413, 3000334414, 3000336487 ***Added 3/21/24*** 3000330091, 3000330092, 3000330094, 3000330095, 3000330096, 3000333095, 3000333096, 3000333097, 3000334409, 3000334410, 3000336488, 3000337834, 3000342904, 3000342905, 3000342906, 3000342907, 3000342908, 3000344446, 3000344447, 3000344471, 3000344472, 3000344474, 3000346048, 3000346049, 3000346050, 3000346051, 3000346052, 3000348543, 3000348546, 3000348552, 3000348553, 3000349793, 3000349795, 3000349796, 3000349797, 3000352614, 3000352615, 3000352616, 3000352617, 3000353439, 3000353440
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1076-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1077-2024 | Class II | Ongoing | Voluntary: Firm initiated |