AFT (Allograft Filler Tube) Straight Tube (1/2 Filled), Product Code 227010 (bone void filler)
Musculoskeletal Transplant Foundation, Inc. began a recall of AFT (Allograft Filler Tube) Straight Tube (1/2 Filled), Product Code 227010 (bone void filler) on . FDA classified it as Class II and gives the reason as "Filling process qualification did not demonstrate adequate process capability. Possible consequences include extended surgery time due to extrudability issues". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1066-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Musculoskeletal Transplant Foundation d/b/a/ MTF Biologics
- Product
- AFT (Allograft Filler Tube) Straight Tube (1/2 Filled), Product Code 227010 (bone void filler)
- Product code
- MQV Filler, Bone Void, Calcium Compound
- Reason for recall
- Filling process qualification did not demonstrate adequate process capability. Possible consequences include extended surgery time due to extrudability issues.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm notified their distributor of the issue during a meeting on December 22, 2023. A notification letter was also issued to the distributor. If the device/tissue was transplanted, no further action is warranted. However, if the device/tissue is still available in the customer's inventory, the unit shall be returned to Spineology, the distributor. Should you have any questions, please contact the firm at 732-661-2209, or email MTF Regulatory Affairs at [email protected].
- Quantity in commerce
- 89
- Distribution
- Product was distributed to Minnesota.
- Product codes and lots
- Donor Lot 02921115411127. UDI-DI: W4184227010T0473
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1066-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1067-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.