Z-1175-2024Class II

Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella...

Abiomed, Inc. / initiated 2021-03-01 / Terminated / root cause: software

Abiomed, Inc. began a recall of Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular) on . FDA classified it as Class II and gives the reason as "Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1175-2024
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abiomed, Inc.
Product
Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)
Product code
OZD Temporary Non-Roller Type Left Heart Support Blood Pump
Reason for recall
Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
Root cause tag
software
FDA root cause
Software design
Action
This is a retrospective report for the March 2021 Technical Bulletin, which described a Product Update; specifically the AIC Version 8.5 Software Update. The software update began rollout on March 4, 2021. According to the firm, Version 8.5 of the software contained bug fixes for the issues seen in version 8.4. The technical bulletin was posted to HeartRecovery.com on March 5, 2021.
Quantity in commerce
4,463 (US)
Distribution
US Nationwide distribution.
Product codes and lots
Product Code 0042-0000-US UDI-DI 00813502010022; Product Code 0042-0010-US UDI-DI 00813502010985; Product Code 0042-0040-US UDI-DI 00813502011401; Software Versions V8.4 and V8.4.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1175-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1175-2024Class IITerminatedVoluntary: Firm initiated