Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella...
Abiomed, Inc. began a recall of Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular) on . FDA classified it as Class II and gives the reason as "Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1175-2024
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abiomed, Inc.
- Product
- Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular Support Systems (cardiovascular)
- Product code
- OZD Temporary Non-Roller Type Left Heart Support Blood Pump
- Reason for recall
- Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- This is a retrospective report for the March 2021 Technical Bulletin, which described a Product Update; specifically the AIC Version 8.5 Software Update. The software update began rollout on March 4, 2021. According to the firm, Version 8.5 of the software contained bug fixes for the issues seen in version 8.4. The technical bulletin was posted to HeartRecovery.com on March 5, 2021.
- Quantity in commerce
- 4,463 (US)
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Product Code 0042-0000-US UDI-DI 00813502010022; Product Code 0042-0010-US UDI-DI 00813502010985; Product Code 0042-0040-US UDI-DI 00813502011401; Software Versions V8.4 and V8.4.1