Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1185-2024Class II

Schiff Reagent, Hotchkiss-McManus, packaged in an amber glass bottle with a black polypropylene cap

EMD Millipore Corporation / initiated 2024-01-26 / Open, Classified / root cause: manufacturing process

EMD Millipore Corporation began a recall of Schiff Reagent, Hotchkiss-McManus, packaged in an amber glass bottle with a black polypropylene cap on . FDA classified it as Class II and gives the reason as "This IVD is colorless. A customer complaint was received noting it appeared pink. A review of inventory confirmed additional stock appeared pink". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1185-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
EMD Millipore Corporation
Product
Schiff Reagent, Hotchkiss-McManus, packaged in an amber glass bottle with a black polypropylene cap.
Product code
HZT Reagent, Schiff
Reason for recall
This IVD is colorless. A customer complaint was received noting it appeared pink. A review of inventory confirmed additional stock appeared pink.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
An URGENT MEDICAL DEVICE RECALL NOTIFICATION LETTER, DATED 26 January 2024 was sent to customers. Required/Recommended Action Please immediately examine your inventory to determine if you have any remaining impacted product and forward a copy of this letter to the personnel in your organization that utilize this product and/or need to be made aware of this issue. Please discard this material and request a refund. Please complete attachment 1 form acknowledging receipt of this letter and return to the noted email address. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this issue. Your notification to customers should include a copy of this notification. You may contact Technical or Customer Service to discuss an appropriate solution for your needs at 1-800-221-1975 or 1-800-325-3010 or 1-800-645-5476. If you would like more information about this customer notification, please contact your local representative or use the link www.sigmaaldrich.com/offices for our website to obtain the phone number and/or email address of your local office.
Quantity in commerce
106 Units (25 US, 81 OUS)
Distribution
Domestic: AL, IL, MA, MN, NJ, PA, TX, VA. Foreign: Canada.
Product codes and lots
Part number: 6073-71, 500mL; Lot Numbers (Expiration Date): 3094 (2025-08-04), 3096 (2025-06-28);

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1185-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1185-2024Class IIOngoingVoluntary: Firm initiated