Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1143-2024Class II

Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement

Voco GmbH / initiated 2024-01-25 / Open, Classified / root cause: manufacturing process

Voco GmbH began a recall of Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement on . FDA classified it as Class II and gives the reason as "Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1143-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Voco GmbH
Product
Ionolux, Catalog (REF) Numbers: a) 2115, b)2117, c) 2118, d) 2119, e) 2120, f) 2122; dental cement
Product code
EMA Cement, Dental
Reason for recall
Due to an error in the manufacturing process, it is possible that an incorrect plunger has been fitted to some capsules. This plunger does not contain the correct reactive component for Ionolux. When these capsules are activated, curing is significantly delayed and incomplete. Processing according to the instructions for use is not possible.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Voco issued a field safety notice to its consignees on 01/25/2024 via email. The notice explained the problem, risk, and requested the product be returned. Those who have sold the product were directed to notify their customers and collect the affected product. For questions contact Mr. Michael Cyll at [email protected].
Quantity in commerce
1021 units
Distribution
Worldwide distribution: US distribution to states of: IA, NY, PA, WA; and OUS (foreign) to countries of: Belgium, Bulgaria, Denmark, Germany, France, Greece, Great Britain, India, Italy, Canada, Quatar, Lithuania, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic, Australia, Saudi Arabia, & Ireland.
Product codes and lots
a) 2115, UDI/DI +E22121151, Lot Numbers: 2346779, 2401515; b) 2117, UDI/DI +E22121171, Lot Numbers: 2347832; c) 2118, UDI/DI +E22121181, Lot Numbers: 2349536; d) 2119, UDI/DI +E22121191, Lot Numbers: 2348501; e) 2120, UDI/DI +E22121201, Lot Numbers: 2350638; f) 2122, UDI/DI +E22121221, Lot Numbers: 2349527

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1143-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1142-2024Class IIOngoingVoluntary: Firm initiated
Z-1143-2024Class IIOngoingVoluntary: Firm initiated