Z-1254-2024Class II
Phantom TTC Nail, 10.0 X 200mm, Right
Paragon 28 / initiated 2023-12-21 / Open, Classified / root cause: sterility
Paragon 28, Inc. began a recall of Phantom TTC Nail, 10.0 X 200mm, Right on . FDA classified it as Class II and gives the reason as "Medical devices distributed, prior to sterilization". As of , the recall status on record is Open, Classified, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1254-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Paragon 28
- Product
- Phantom TTC Nail, 10.0 X 200mm, Right
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Medical devices distributed, prior to sterilization
- Root cause tag
- sterility
- FDA root cause
- Under Investigation by firm
- Action
- On January 5, 2024, Paragon 28 requested their consignee(s) return the affected device. On February 12, 2024, Paragon 28 followed up with customers by issuing a "Urgent Medical Device Recall" Notification via email. Paragon informed consignees of the following: 1. There is no action needed from your agency as the three affected devices have been returned and quarantined at Paragon 28 headquarters. 2. The information provided to your agency from Paragon 28 Sales Support related to the sterilization of these devices was provided in error and is not to be followed as other methods for sterilization have not been verified/validated. 3. Any need for this product should be directed to [email protected]. 4. Paragon 28, Inc. is actively working on a solution to this failure under a corrective action investigation. 5. Please complete the attached Recall Response Form to acknowledge this letter.
- Distribution
- US:CA OUS: None
- Product codes and lots
- PA23005925
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1254-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1255-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1256-2024 | Class II | Ongoing | Voluntary: Firm initiated |