Z-1225-2024Class II

VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)

Ortho-Clinical Diagnostics, Inc. / initiated 2024-01-26 / Open, Classified / root cause: software

Ortho-Clinical Diagnostics, Inc. began a recall of VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished) on . FDA classified it as Class II and gives the reason as "A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1225-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Root cause tag
software
FDA root cause
Software Design Change
Action
On January 26, 2024, the firm sent a customer letter to all customers who have V3.8.1 installed on their VITROS Analyzer. Customers were informed of this issue and instructed to follow the instructions when using the VITROS System onboard QC package to update the QC baseline. The software will be updated to mitigate both issues within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
Quantity in commerce
4 US; 18 OUS
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
Product codes and lots
VITROS System Software Version 3.8.1 Regular UDI-DI 10758750012343 Refurbished UD-DI 10758750033201 Serial Numbers US: J46000240 J46000898 J46001030 J46001293 OUS: J46000150 J46001041 J46000190 J46001042 J46000499 J46001186 J46000564 J46001187 J46000801 J46001210 J46000821 J46001217 J46000910 J46001289 J46000933 J46001639 J46001029 J46001645

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1225-2024