Z-1227-2024Class II

VITROS XT 3400 Chemistry System, Catalog No. 6844458

Ortho-Clinical Diagnostics, Inc. / initiated 2024-01-26 / Open, Classified / root cause: software

Ortho-Clinical Diagnostics, Inc. began a recall of VITROS XT 3400 Chemistry System, Catalog No. 6844458 on . FDA classified it as Class II and gives the reason as "A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1227-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS XT 3400 Chemistry System, Catalog No. 6844458
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.
Root cause tag
software
FDA root cause
Software Design Change
Action
On January 26, 2024, the firm sent a customer letter to all customers who have V3.8.1 installed on their VITROS Analyzer. Customers were informed of this issue and instructed to follow the instructions when using the VITROS System onboard QC package to update the QC baseline. The software will be updated to mitigate both issues within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
Quantity in commerce
7 US; 4 OUS
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
Product codes and lots
VITROS System Software Version 3.8.1 UDI-DI 10758750031986 Serial Numbers US: J34500180 J34500289 J34500422 J34500473 J34500622 J34500861 J34500887 OUS: J34500515 J34500609 J34500802 J34500940

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1227-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1224-2024Class IIOngoingVoluntary: Firm initiated
Z-1225-2024Class IIOngoingVoluntary: Firm initiated
Z-1226-2024Class IIOngoingVoluntary: Firm initiated
Z-1227-2024Class IIOngoingVoluntary: Firm initiated
Z-1228-2024Class IIOngoingVoluntary: Firm initiated