BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352
CareFusion 303, Inc. began a recall of BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352 on . FDA classified it as Class II and gives the reason as "Automated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or crash, which may result in delays in medication access, and potential...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1223-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- CareFusion 303, Inc
- Product
- BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352;
- Product code
- BRY Cabinet, Table And Tray, Anesthesia
- Reason for recall
- Automated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or crash, which may result in delays in medication access, and potential data loss.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 2/1/24, correction notices were mailed or emailed to customers and distributors who were told the following: Devices with a random application hang, freeze and/or crashing can be resolved with a reboot of the device. In downtime scenarios, medications can be obtained from other devices or from pharmacy. There are also keys that can be used to manually access contents within the device. 1) Ensure you continue to manually reboot your devices as a part of your maintenance schedule. 2. Disseminate this notification within your facility network to ensure they are aware of the issue. 3. If you experience this issue, please contact Technical Support. 4. Complete the Customer Response Form and return to [email protected] If you require further assistance, please contact the firm: Customer Support: 1-866-583-8783, 5:00am PST to 4:00pm PST (8:00am ET to 7:00pm ET), Monday - Friday. Technical Support: bd.com/self-service, 1-800-727-6102, 24 hours, 7 days a week
- Quantity in commerce
- 20,443
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of VA, SC, IL, AL, KS, LA, GA, FL, CA, PR, PA, WV, WA, HI, NC, MD, AR, NY, ND, TX, NM, GU, MS, AZ, DC, AK, NE, DE, CO, SD, MO, NJ, OK, NV, OH, KY, WY, MT, UT, ID, MA, MI, MP, NH, IA, TN, CT, RI, IN, WI, OR, MN, VT, ME and the countries of BH, DE, GB, IT, JP, ES, GU, KR, BE, AR, BR, BS, CL, CO, MX, AE, AU, CA, EG, FR, HK, MY, NZ, QA, SA, TH, NL, MC, CH, FI, JO, KW, LB, OM.
- Product codes and lots
- REF/UDI-DI: 323/10885403512667, 327/10885403477836, 352/10885403512674. All Using Software Version ES System Release 1.4.x, 1.5.x and 1.6.x
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1223-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.