Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1223-2024Class II

BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352

CareFusion 303, Inc / initiated 2024-02-01 / Open, Classified / root cause: software

CareFusion 303, Inc. began a recall of BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352 on . FDA classified it as Class II and gives the reason as "Automated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or crash, which may result in delays in medication access, and potential...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1223-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352;
Product code
BRY Cabinet, Table And Tray, Anesthesia
Reason for recall
Automated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or crash, which may result in delays in medication access, and potential data loss.
Root cause tag
software
FDA root cause
Software design
Action
On 2/1/24, correction notices were mailed or emailed to customers and distributors who were told the following: Devices with a random application hang, freeze and/or crashing can be resolved with a reboot of the device. In downtime scenarios, medications can be obtained from other devices or from pharmacy. There are also keys that can be used to manually access contents within the device. 1) Ensure you continue to manually reboot your devices as a part of your maintenance schedule. 2. Disseminate this notification within your facility network to ensure they are aware of the issue. 3. If you experience this issue, please contact Technical Support. 4. Complete the Customer Response Form and return to [email protected] If you require further assistance, please contact the firm: Customer Support: 1-866-583-8783, 5:00am PST to 4:00pm PST (8:00am ET to 7:00pm ET), Monday - Friday. Technical Support: bd.com/self-service, 1-800-727-6102, 24 hours, 7 days a week
Quantity in commerce
20,443
Distribution
Worldwide distribution - US Nationwide distribution including in the states of VA, SC, IL, AL, KS, LA, GA, FL, CA, PR, PA, WV, WA, HI, NC, MD, AR, NY, ND, TX, NM, GU, MS, AZ, DC, AK, NE, DE, CO, SD, MO, NJ, OK, NV, OH, KY, WY, MT, UT, ID, MA, MI, MP, NH, IA, TN, CT, RI, IN, WI, OR, MN, VT, ME and the countries of BH, DE, GB, IT, JP, ES, GU, KR, BE, AR, BR, BS, CL, CO, MX, AE, AU, CA, EG, FR, HK, MY, NZ, QA, SA, TH, NL, MC, CH, FI, JO, KW, LB, OM.
Product codes and lots
REF/UDI-DI: 323/10885403512667, 327/10885403477836, 352/10885403512674. All Using Software Version ES System Release 1.4.x, 1.5.x and 1.6.x

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1223-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1223-2024Class IIOngoingVoluntary: Firm initiated