Z-1279-2024Class II
D-Clip Standard/Mini Applier, 230mm bayonet, jaw rigid, double action
Peter Lazic GmbH / initiated 2023-10-06 / Open, Classified / root cause: design
Peter Lazic Gmbh began a recall of D-Clip Standard/Mini Applier, 230mm bayonet, jaw rigid, double action on . FDA classified it as Class II and gives the reason as "Customer complaint received that reported the applier failed during surgery. The applier did not release the clip. The applier remained attached to the clip". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1279-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Peter Lazic GmbH
- Product
- D-Clip Standard/Mini Applier, 230mm bayonet, jaw rigid, double action.
- Product code
- HCI Applier, Aneurysm Clip
- Reason for recall
- Customer complaint received that reported the applier failed during surgery. The applier did not release the clip. The applier remained attached to the clip.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Reportable Event - Removal letters dated October 6, 2023 were sent to customers. Customers were instructed to return all product subject to recall and to notify any of their customers as well. Please return items for the attention of one of the PRRCs provided to ensure quick processing and assignment to the case. If you have any questions, please contact Mr. Sven Lazic ([email protected] +49 (0) 7461 96643 - 20) or Ms. Kim Maier ([email protected] or +49 (0) 7461 96643 - 21).
- Quantity in commerce
- 10 units
- Distribution
- US Nationwide distribution in the state of ME.
- Product codes and lots
- Model number: 35.412; UDI/DI: 04250603739616; All Lot numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1277-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1278-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1279-2024 | Class II | Ongoing | Voluntary: Firm initiated |