Z-1293-2024Class II
EasyPoint Needle, 25G x 5/8", REF: 82091
Retractable Technologies, Inc. / initiated 2024-02-05 / Open, Classified / root cause: manufacturing process
Retractable Technologies, Inc. began a recall of EasyPoint Needle, 25G x 5/8", REF: 82091 on . FDA classified it as Class II and gives the reason as "The needle cannula of a retractable needle may detach from the needle holder in the patient". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1293-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Retractable Technologies, Inc.
- Product
- EasyPoint Needle, 25G x 5/8", REF: 82091
- Product code
- FMI Needle, Hypodermic, Single Lumen
- Reason for recall
- The needle cannula of a retractable needle may detach from the needle holder in the patient.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 2/5/24, recall notices were sent to customers and distributors who were asked to do the following: 1) Segregate, quarantine, and discontinue use and distribution of affected devices. 2) Inform all personnel who have received or are receiving affected devices that they should not use affected devices. 3) If affected product was further distributed, notify these additional customers immediately. When notifying customers include a copy of the recall notice, label images, and the response form. 4) Complete and return the response form to [email protected] Customers with additional questions can email the firm at [email protected]
- Quantity in commerce
- 477,600
- Distribution
- US Nationwide distribution including in the states of NY, FL, IL, MA, UT, NJ, AZ, MO, WA, NC, CA, TX, NV, IN, PA, VA, IA, GA, TN, MI, OH, OK, CO, OR, SC, MN, CT.
- Product codes and lots
- UDI-DI: 00613703820907. Lot: K220402. Expiration: 03/28/2027
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1293-2024 | Class II | Ongoing | Voluntary: Firm initiated |