Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1390-2024Class II

Boston Scientific iSLEEVE EXPANDABLE Introducer Set REF H74939349140, facilitates femoral access to the...

Boston Scientific Corporation / initiated 2024-02-20 / Open, Classified / root cause: manufacturing process

Boston Scientific Corporation began a recall of Boston Scientific iSLEEVE EXPANDABLE Introducer Set REF H74939349140, facilitates femoral access to the vascular system on . FDA classified it as Class II and gives the reason as "Boston Scientific is conducting a removal of specific batches of the iSLEEVETM 14F Expandable Introducer Set due to complaints associated with a hemostatic valve leak". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1390-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific iSLEEVE EXPANDABLE Introducer Set REF H74939349140, facilitates femoral access to the vascular system
Product code
DYB Introducer, Catheter
Reason for recall
Boston Scientific is conducting a removal of specific batches of the iSLEEVETM 14F Expandable Introducer Set due to complaints associated with a hemostatic valve leak.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Boston Scientific issued an Urgent Field Safety Notice - Medical Device Product Removal to its consignees on 02/20/2027 via FedEx. The notice explained the problem with the device, potential risk, and requested the following: 1. Immediately discontinue use and segregate the affected product. 2. Complete and return the RVTF to get a Return Goods Authorization (RGA) number. 3. Once Boston Scientific receives your completed RVTF, you will be contacted within 2 business days and provided an RGA number for product return. For questions, contact your local Boston Scientific representative.
Quantity in commerce
4060 devices
Distribution
Worldwide distribution - US Nationwide and the countries of Canada.
Product codes and lots
GTIN 08714729950660, Batch Numbers: 31368041, 31390084, 31390088, 31390089, 31501866, 31501867, 31501868, 31501869, 31520897, 31520898, 31604492, 31616143, 31633635, 31650457, 31664216, 31664217, 31671140, 31671143, 31693638, 31693639, 31724952, 31732930, 31732931, 31740806, 31740807, 31740808, 31751118, 31777134, 31786582, 31786583, 31818290, 31818299, 31818341, 31818342, 31818343, 31818344, 31818345, 31818346, 31818347, 31838684, 31838685, 31838686, 31838687, 31838688, 31838689, 31877302, 31877304, 31877305, 31877306, 31904748, 31904749, 31923755, 31923756, 31953345, 31953348, 32008170, 32031191, 32031192, 32031193, 32031196, 32059171, 32059386, 32059387, 32059388, 32059392, 32059393, 32059394, 32059396, 32059397, 32212151, 32212153, 32230544, 32230545, 32230546, 32230547, 32230548, 32230549, 32230550, 32230551, 32284196, 32284198, 32284200, 32304326, 32304327, 32304328, 32313757, 32313758, 32313759, 32322405, 32322406, 32322407, 32331162, 32331163, 32331164, 32331165, 32427040, 32452794, 32461848, 32461849, 32461850, 32470848, 32470849, 32480089, 32480090, 32481344, 32481345, 32522178, 32530450, 32530451, 32602892, 32602893, 32611557, 32611558, 32611559, 32620708, 32620709, 32620710, 32629426, 32642560, 32642561

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1390-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1390-2024Class IIOngoingVoluntary: Firm initiated