Z-1820-2024Class II

HeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450P

HeartSine Technologies Ltd / initiated 2024-04-08 / Open, Classified / root cause: manufacturing process

HeartSine Technologies Ltd began a recall of HeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450P on . FDA classified it as Class II and gives the reason as "Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user during use of the device. Visual instructional icons will still be...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1820-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
HeartSine Technologies Ltd
Product
HeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450P
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user during use of the device. Visual instructional icons will still be present, but this issue could potentially lead to no therapy or a delay in therapy.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 04/08/24 Stryker sent correction notices to customers informing them of the following: 1) Follow instructions to power cycle affected devices: a) Check the expiration date on the Pad-Pak and if passed, do not use and replace the expired Pad-Pak. b) Place the affected device face up on a flat surface and slide the Pad-Pak into the affected device until you hear the "double click." c) Verify that the green Status indicator is blinking to indicate the initial self-test routine has been performed and the device is ready for use. d) Press On/Off button. c) Listen for, but do not follow, voice prompts. -If you do not hear a prompt, contact your Authorized Distributor or HeartSine Technologies Technical Support at: [email protected] -If you hear prompt "Adult patient" and/or "Call for medical assistance", no further action is needed. d) Press On/Off button. If you have not heard a warning message and the status indicator continues to flash green, the device is ready for use. e) Carry out check once every three months and maintain awareness. f) Audio issue will not cause a warning message, but if any other warning messages are played, or you see a red flashing status indicator, please refer to User Manual (General Troubleshooting). 2) Complete/return the response form to [email protected]. 3) If further distributed, email to [email protected]. Stryker will work with you to ensure recipients are notified. 4) If you have any questions or concerns, contact Customer Support at +44 28 9093 9400, 9A.M. to 5P.M. (GMT), or by email at [email protected]. On 3/17/2025, additional correction notices were sent and in addition customers were informed: - Keep track of the total number of impacted devices as well as the number of devices that do not provide audio prompts (if any). - If unable to complete this check every 3 months, please reach out to [email protected] for further alternative solutions that may be applied.
Quantity in commerce
183,248
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, GC, HI, IA, ID, IL, IN, KS, KY, LA, MA, MB, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY.
Product codes and lots
UDI-DI: 5060167125379, 5060167120688, 5060167125720, 5060167128066, 5060167122330, 5060167122491, M727SAM450P, 5060167122514, 5060167121371, 5060167120671, 5060167121630, 5060167128042, 5060167120701, 5060167128080, 5060167120695, 5060167120725, 5060167120718, 5060167120909, 5060167120916, 5060167122071, 5060167120893. Devices affected by this notification begin with the following prefixes and device codes: 16D, 16E, 16G, 17D, 17E, 17G, 18D, 18E, 18G, 19D, 19E, 19G, 20D, 20E, 20G, 21D, 21E, 21G, 22D, 22E, 22G, 23D, 23E, 23G, 24D, 24E, 24G Serial Numbers: The prefix (device identifier) consists of the manufacturing date (YY) and the device model (B, D, E, G, or H) and an 8-digit serial number string. Example: 16D00001234

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1820-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1820-2024Class IIOngoingVoluntary: Firm initiated