Z-1326-2024Class I

stay-safe /Luer Lock Catheter Ext. 18 in

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2024-01-23 / Open, Classified / root cause: design

Fresenius Medical Care Holdings, Inc. began a recall of stay-safe /Luer Lock Catheter Ext. 18 in on . FDA classified it as Class I and gives the reason as "The peroxide cross-linked tubing has leachables identified as NDL PCBAs". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1326-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
stay-safe /Luer Lock Catheter Ext. 18 in
Product code
KDJ Set, Administration, For Peritoneal Dialysis, Disposable
Reason for recall
The peroxide cross-linked tubing has leachables identified as NDL PCBAs.
Root cause tag
design
FDA root cause
Device Design
Action
IMPORTANT MEDICAL DEVICE CORRECTION Letter were sent to customer on February 14, 2024 informing customers that non-dioxin-like polychlorinated biphenyl acids (PCBAs) could leach from the stay-safe Catheter Extension sets and Adapter, which are manufactured using peroxide cross-linked silicone tubing. Please note that PCBAs are chemically different from polychlorinated biphenyls (PCBs) which were not identified. No complaints have been reported. PCBA levels decrease with shorter catheter extension sets and decrease over the treatment time. PCBA exposure threshold examples were provided per body weight (less than 10kg and between 10 and 40 kg), treatment length and various device combinations. Patients with body weight greater than 40kg are not impacted. FMCRTG recommends the following for Healthcare providers until an update to the catheter extension silicone material becomes available: 1. Use only the shortest length extension set (6-inch, Part No. 050-95013) when treating patients with a body weight below 40 kg, including infants and neonates. 2. Remain alert for further updates and recommendations from the FMCRTG and FDA. 3. Please report any complaints or adverse events to [email protected] or FDA MedWatch at https://www.fda.gov/safety/medical-product-safetyinformation/medwatch-forms-fda-safety-reporting. For any clinical or medical questions, please contact the Global Medical Information and Education Office at 1-855-616-2309. ***Update 04/10/2024*** A revised Medical Device Correction Letter and Poster were sent to customers via email recommending the following for Healthcare providers until an update to the catheter extension silicone material becomes available: 1. Use only the shortest length extension set (6-inch, Part No. 050-95013) when treating patients with a body weight below 40 kg, including infants and Neonates and if possible, avoid using the adapter. This is for both existing patients and patients about to initiate peritoneal dialysis. Re
Quantity in commerce
479,978 eaches
Distribution
Domestic: Nationwide Distribution.
Product codes and lots
Model Number: 050-95005; UDI/DI (Bag): 00840861100798; UDI/DI (Case): 10840861100795; All lots.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1326-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1323-2024Class IOngoingVoluntary: Firm initiated
Z-1324-2024Class IOngoingVoluntary: Firm initiated
Z-1325-2024Class IOngoingVoluntary: Firm initiated
Z-1326-2024Class IOngoingVoluntary: Firm initiated
Z-1327-2024Class IOngoingVoluntary: Firm initiated