Z-1369-2024Class II

Cystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay on...

Randox Laboratories, Limited / initiated 2024-02-14 / Open, Classified / root cause: manufacturing process

Randox Laboratories Ltd. began a recall of Cystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay on automated analysers Catalog Number: CVS 2699 on . FDA classified it as Class II and gives the reason as "Cystatin C Calibrator Series CVS 2699, is running with a negative bias compared to other methods". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1369-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Randox Laboratories, Limited
Product
Cystatin C Calibrator - In vitro diagnostic (IVD) use in the calibration of the Randox Cystatin C assay on automated analysers Catalog Number: CVS 2699
Product code
NDY Test, Cystatin C
Reason for recall
Cystatin C Calibrator Series CVS 2699, is running with a negative bias compared to other methods
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Randox Laboratories Ltd, UK ( manufacturer) notified the US Distributor Randox Laboratories USA . The distributor initial contact and the follow up contact via email on 2/20/24. Letter states reason for recall, health risk and action to take: We have realigned the target values in the above lot numbers to the IFCC Reference Standard. Please refer to the table below for the updated calibrator targets. You may experience a positive shift in Quality Control and patient sample recovery. If you are using CVS 2699 batch 600957 lots 1378CY-1382CY please provide proof of scrappage and a replacement calibrator lot will be provided. Please discard all copies of the calibrator Instructions For Use (IFU)s and download the updated IFUs from www.randox.com. If further information is required, please contact [email protected] . In line with the realignment to the Cystatin C Calibrator Series, we are updating the target and range for the Cystatin C controls, CYS5019 and CYS5020. The updated target and range can be found in the table below. Please discard any copies of the IFU and download the updated version from www.randox.com.
Quantity in commerce
666 units
Distribution
US Nationwide distribution in the states of CA, IN.
Product codes and lots
GTIN: 05055273207439 (1) Batch 600957 Lot 1378CY - 1382CY (2) Batch 621809, 621815 Lot 1403CY - 1407CY (3) Batch 650533 Lot 1414CY - 1418CY

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1369-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1369-2024Class IIOngoingVoluntary: Firm initiated