Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1370-2024Class II

SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs...

Alcon Laboratories / initiated 2024-02-21 / Open, Classified / root cause: manufacturing process

Alcon Research, LLC began a recall of SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44, 5675-76, 13554-11... on . FDA classified it as Class II and gives the reason as "Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1370-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Alcon Laboratories
Product
SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44, 5675-76, 13554-11, 18018-06, 15190-16 , 15190-16, 15190-16, 15190-16, 15190-16, 13554-11, 5918-39, 15190-16, 15190-16, 9208-66, 9208-66, 5675-76, 5918-39, 9208-66, 15190-16, 9065-25, 5918-39, 16066-15, 16066-15
Product code
LRO General Surgery Tray
Reason for recall
Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so after the packs are opened, syringes should be discarded at surgical case end, and not reshelved, because the use of non-sterile bulb irrigation syringes may increase infection risk.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 2/21/24, correction notices were mailed to customers who were asked to do the following: 1) Review your inventory to determine if you have any affected product within your facility. 2) If you have any affected Custom Pak lots in inventory, please use the included stickers to label any affected product. If you have previously reshelved either of the affected Medline Bulb Irrigation Syringe lots, dispose them. 3) All components inside the Pak are intended to be used only for one specific surgical case and should be discarded at the end of the surgical case. 4) Forward the correction notice to all departments within your organization who may be in possession of affected product; and any other organization to which this product may have been transferred. 5) Complete and return the response form via email to [email protected] Customers with questions or concerns can call the firm at 1-800-862-5266.
Quantity in commerce
886
Distribution
US Nationwide distribution in the states of OH, WV.
Product codes and lots
Product/Lot/Expiration/UDI-DI: 18018-06/15T707/02/28/2025/380655148440, 3865-44/15T2RC/02/28/2025/380655148150, 5675-76/15T70E/02/28/2025/380655148174, 13554-11/15V4RC/02/28/2025/380655143711, 18018-06/15V3E0/02/28/2025/380655148440, 15190-16/JZ5179517/06/30/2024/380655148303, 15190-16/JZ5179519/06/30/2024/380655148303, 15190-16/JZ5179521/06/30/2024/380655148303, 15190-16/JZ5179523/06/30/2024/380655148303, 15190-16/JZ5179529/06/30/2024/380655148303, 13554-11/15TAUD/01/31/2026/380655143711, 5918-39/15TE5F/02/28/2025/380655149140, 15190-16/15THVA/04/30/2025/380655148303, 15190-16/15TNXJ/04/30/2025/380655148303, 9208-66/15V8CU/04/30/2025/380651693326, 9208-66/15XVDD/05/31/2025/380651693326, 5675-76/JZ5180811/06/30/2024/380655148174, 5918-39/JZ5180812/06/30/2024/380655149140, 9208-66/JZ5180332/03/31/2025/380651693326, 15190-16/JZ5180856/05/31/2025/380655148303, 9065-25/JZ5187593/11/30/2025/380651682849, 5918-39/JZ5187260/06/30/2024/380655149140, 16066-15/JZ5190913/07/31/2025/380655149683, 16066-15/JZ5191891/07/31/2025/380655149683 Bulb Syringe lots: 96922110001 or 96923100001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1370-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1370-2024Class IIOngoingVoluntary: Firm initiated