Z-1350-2024Class II

Convatec EsteemBody Soft Convex, Drainable Pouch, Product codes (REF): a) 423643, b) 423658, c) 423659, d)...

ConvaTec Limited / initiated 2024-02-19 / Open, Classified / root cause: packaging

ConvaTec, Inc began a recall of Convatec EsteemBody Soft Convex, Drainable Pouch, Product codes (REF): a) 423643, b) 423658, c) 423659, d) 423664, e) 423666, f) 423645, g) 423667, h) 423657... on . FDA classified it as Class II and gives the reason as "Convatec Inc is conducting a Voluntary Medical Device Recall (removal) for specific lots of the EsteemBody Soft Convex, Drainable Pouch due to inconsistency in the strength of the filter weld". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1350-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
ConvaTec Limited
Product
Convatec EsteemBody Soft Convex, Drainable Pouch, Product codes (REF): a) 423643, b) 423658, c) 423659, d) 423664, e) 423666, f) 423645, g) 423667, h) 423657, i) 423652, j) 423646, k) 423654, l) 423653, m) 423647, n) 423644, o) 423648, p) 423665, q) 423656, r) 423651, s) 423660, t) 423661
Product code
FON Bag, Drainage, With Adhesive, Ostomy
Reason for recall
Convatec Inc is conducting a Voluntary Medical Device Recall (removal) for specific lots of the EsteemBody Soft Convex, Drainable Pouch due to inconsistency in the strength of the filter weld.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
ConvaTec issued an URGENT: VOLUNTARY MEDICAL DEVICE RECALL notice to its consignees on 02/19/2024 via email. The notice explained the problem and requested the consignee quarantine the affected product and respond. Upon receipt, the firm will issue a Certificate of destruction and request the destruction of the product. The notice be shared with the correct individual with the organization and disseminated to all who need to be aware and distributors were directed to notify their customers.
Quantity in commerce
6709 units
Distribution
Worldwide distribution - US Nationwide and the countries of Italy, UK.
Product codes and lots
a) 423643, UDI/DI 768455203925, Lot Numbers 3L03391; b) 423658, UDI/DI 768455204076, Lot Numbers 3L04005; c) 423659, UDI/DI 768455204083, Lot Numbers 3L03670; d) 423664, UDI/DI 768455204137, Lot Numbers 3L02230; e) 423666, UDI/DI 768455204151, Lot Numbers 3L04004; f) 423645, UDI/DI 976845520394, Lot Numbers 3L04989; g) 423667, UDI/DI 768455204168, Lot Numbers 3L04674; h) 423657, UDI/DI 768455204069, Lot Numbers 3L04114; i) 423652, UDI/DI 768455204014, Lot Numbers 3L03669; j) 423646, UDI/DI 768455203956, Lot Numbers 3M01542; k) 423654, UDI/DI 768455204038, Lot Numbers 3M00021; l) 423653, UDI/DI 768455204021, Lot Numbers 3M00786; m) 423647, UDI/DI 768455203963, Lot Numbers 3M01540; n) 423644, UDI/DI 768455203932, Lot Numbers 3M01539; o) 423648, UDI/DI 768455203970, Lot Numbers 3M01541; p) 423665, UDI/DI 768455204144, Lot Numbers 3M00029; q) 423656, UDI/DI 768455204052, Lot Numbers 3L00958; r) 423651, UDI/DI 768455204007, Lot Numbers 3K03888; s) 423660, UDI/DI 768455204090, Lot Numbers 3M00022; t) 423661, UDI/DI 768455204106, Lot Numbers 3M00023

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1350-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1350-2024Class IIOngoingVoluntary: Firm initiated