Nimbus Administration Set; Nimbus Flex; Nimbus PainPro (a.k.a. Halo Ambulatory Infusion System); Catalog...

InfuTronix LLC / initiated 2024-02-21 / Open, Classified / root cause: design

InfuTronix LLC began a recall of Nimbus Administration Set; Nimbus Flex; Nimbus PainPro (a.k.a. Halo Ambulatory Infusion System); Catalog Number IT1025. Ambulatory Infusion Pump and components on . FDA classified it as Class I and gives the reason as "InfuTronix is removing the Nimbus Infusion Pump Systems from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1285-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
InfuTronix LLC
Product
Nimbus Administration Set; Nimbus Flex; Nimbus PainPro (a.k.a. Halo Ambulatory Infusion System); Catalog Number IT1025. Ambulatory Infusion Pump and components
Product code
FRN Pump, Infusion
Reason for recall
InfuTronix is removing the Nimbus Infusion Pump Systems from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump housing. The device will not be available or supported after June 20, 2024.
Root cause tag
design
FDA root cause
Device Design
Action
On February 21, 2024, the firm began notifying customers via URGENT MEDICAL DEVICE REMOVAL letters. Customers were informed that the identified Nimbus Infusion Pumps are being removed from the market. The products will no longer be supported by InfuTronix for the Nimbus Infusion Pumps or related infusion sets beyond June 20, 2024. Customers of InfuTronix who have Nimbus Infusion Pumps and associated infusion sets in their possession should contact InfuTronix customer service at [email protected], where they will be instructed on the Return Material Authorization (RMA) process to be followed for returning the both the Nimbus Infusion Pumps and associated infusion sets. As the device will not be available or supported after June 20, 2024, InfuTronix encourages the healthcare provider to seek alternative methods of drug product infusion appropriate for their patient s needs, based on their medical expertise, at the earliest possible opportunity. ***Updated 4/12/24*** The firm issued updated letters on 4/9/24 to reflect that the recall had been classified Class I by the FDA. Effective immediately, Nimbus Pumps will no longer be shipped. However, administration sets will continue to be shipped through June 20, 2024.
Quantity in commerce
52,328 total
Distribution
Domestic US distribution nationwide. No international distribution.
Product codes and lots
UDI-DI 00817170020000 Lots A100-150205200 A100-150415015 A100-150430270 A100-150611300 A100-150716620 A100D-1705191000 AZ060-171201564 AZ060-1801021020 AZ060-180609264 AZ060-180626552 AZ060-190104792 AZ060-1902151200 A100F-160707073 A100F-160908250 A100F-161114700 All units recalled

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1285-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1285-2024Class IOngoingVoluntary: Firm initiated
Z-1286-2024Class IOngoingVoluntary: Firm initiated