Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1469-2024Class II

Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTS

Medtronic, Inc. / initiated 2024-02-28 / Open, Classified / root cause: labeling and UDI

Covidien, LP began a recall of Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTS on . FDA classified it as Class II and gives the reason as "The potential for trocar seal disengagement when using mesh products incorrectly with the device". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1469-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTS
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
The potential for trocar seal disengagement when using mesh products incorrectly with the device.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On February 28, 2024, the firm notified customers of the issue via URGENT: MEDICAL DEVICE RECALL letters. Actions to be taken by the customer: 1. Inform all surgeons and clinicians who handle the preparation and/ or placement of a Mesh Device that utilize the balloon and blunt tip trocar devices. 2. Prior to using any Mesh device in conjunction with the following Trocars: Covidien Auto Suture" Structural Balloon Trocar and Auto Suture" Blunt Tip Trocar, please carefully review and adhere to the mesh manufacturer s Instructions for Use (IFU) on proper insertion techniques. 3. Please complete and return the customer confirmation form enclosed with this letter acknowledging receipt of this information via email to [email protected]. 4. Please transfer this notice to other organizations on which this action has an impact and maintain a copy of this notice in your records. Adverse events or quality problems experienced with this product should be reported to the FDA and Medtronic: Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm, or call FDA at 1-800-FDA-1088 (1-800-332-1088) and Medtronic Vascular Customer Service at 800-716-6700. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Representative.
Quantity in commerce
18,048 units
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Aruba, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic Of, Luxembourg, Macao, Malaysia, Malta, Martinique, Mayotte, Mexico, Netherlands, New Caledonia, New Zealand, Nigeria, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, United States, Uzbekistan, Viet Nam, Virgin Islands, U.S.;
Product codes and lots
Product Number/CFN: OMS-T10BTS; UDI/DI: 20884521081076, 10884521081079; Lot Serial Number: P1D1416, P1E1509, P1F1547, P1H0898, P1J0874, P1J0895, P1J0895, P1J1268, P1J1268, P1J1340, P1J1340, P1J1341, P1K0181, P1L0683, P1L0978, P1L0984, P1L0984, P2A0017, P2A0017, P2A0018, P2A0019, P2A0019, P2A0020, P2A0021, P2A0021, P2A0750, P2A0750, P2B0135, P2B0136, P2B0136, P2B0752, P2B0916, P2C0757, P2D0347, P2D0541, P2D0542, P2F0385, P2F0386, P2F0386, P2H0022, P2H0022, P2H0147, P2H0147, P2K0433, P2K0433, P2K0434, P2K0434, P3D0123, P3D0123, P3D0124, P3D0124, P3E0187, P3E0187, P3E0188, P3E0189, P3E0189, P3H1197, P3J0771, P3J0858, P3K1341;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1469-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1465-2024Class IIOngoingVoluntary: Firm initiated
Z-1466-2024Class IIOngoingVoluntary: Firm initiated
Z-1467-2024Class IIOngoingVoluntary: Firm initiated
Z-1468-2024Class IIOngoingVoluntary: Firm initiated
Z-1469-2024Class IIOngoingVoluntary: Firm initiated
Z-1470-2024Class IIOngoingVoluntary: Firm initiated
Z-1471-2024Class IIOngoingVoluntary: Firm initiated