Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Randox Laboratories Ltd. began a recall of Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567 on . FDA classified it as Class II and gives the reason as "Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1593-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Randox Laboratories, Limited
- Product
- Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
- Product code
- JIT Calibrator, Secondary
- Reason for recall
- Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- Randox Laboratories (Manufacturer) issued Urgent Field Safety Notice on 31 Jan 24 Complaint Reference: REC716 letter via email to Distributors. Distributors will contact the end-users. Letter states reason for recall, health risk and action to take: We have reassigned the target values in the above lot numbers in line with the ERM reference material. Please refer to the table below for the updated calibrator targets. You may experience a shift in Quality Control and patient sample recovery of up to +12%. Please discard all copies of the calibrator Instructions For Use (IFU)s and download the updated IFUs from www.randox.com . If further information is required, please contact [email protected] . In line with the restandardisation to the Microalbumin Calibrator Series, we are updating the target and range for the Microalbumin Liquid control, MA1361, lots 1102MA and 1103MA which are present in batch 631698. The updated targets and ranges can be found in the table below. Please discard any copies of the IFU and download the updated version from www.randox.com.
- Quantity in commerce
- 8 units
- Distribution
- US Nationwide distribution in the states of GA, MA, WV including PR.
- Product codes and lots
- GTIN: 05055273204520 (1) Batch/Lot: 619282 / 1094MA-1098MA
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1593-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1594-2024 | Class II | Ongoing | Voluntary: Firm initiated |