Z-1476-2024Class II

ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20

Abbott Medical / initiated 2024-02-21 / Open, Classified / root cause: manufacturing process

Abbott Laboratories began a recall of ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20 on . FDA classified it as Class II and gives the reason as "The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1476-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
ARCHITECT STAT Myoglobin Reagent Kit, LN 2K43-25 and 2K43-20.
Product code
DDR Myoglobin, Antigen, Antiserum, Control
Reason for recall
The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis of myocardial infarction.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recalling firm issued letters dated 2/21/2024 on the same date via FedEx Priority Overnight Express. The letter explained the reason for recall, the impact on patient results, and the necessary actions to be taken by the customer. Those actions included immediately discontinuing the use of and destroy any remaining inventory of the affected lots according to the consignee's laboratory procedures; review the letter with the Medical Director or Laboratory Management and follow the laboratory protocol regarding the need for reviewing previously reported results generated with lot 50808UN23; immediately contact Customer Support for replacement material; complete and return the Customer Reply Form; and if the product has been forwarded to other laboratories, they are to be informed of the recall with a copy of this letter provided to them. The Customer Reply Form was to be returned to the recalling firm prior to 06MAR2024.
Quantity in commerce
7482 kits
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IL, KS, LA, MI, MO, MS, NE, NY, OH, OR, PA, TX, VA, and WI. The countries of Angola, Austria, Belarus, Belgium, Brazil, Czech Republic, France, Germany, Hungary, India, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Luxembourg, Mali, Mexico, Morocco, Poland, Portugal, Peoples Republic of China, Romania, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Switzerland, Taiwan, Tajikistan, Tanzania, Trinidad & Tobago, Turkey, Uganda, Uruguay, Uzbekistan, and Vietnam.
Product codes and lots
LN 2K43-25 - Lot number 50808UN23, exp. 11/30/2024, UDI (01)00380740003302 (17)241130(10)50808UN23; LN 2K43-20 - Lot number 60104UN23, exp. 11/30/2024, UDI (01)00380740003296 (17)241130(10)60104UN23.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1476-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1476-2024Class IIOngoingVoluntary: Firm initiated