Z-1482-2024Class II

RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A...

RaySearch Laboratories AB (publ) / initiated 2024-02-23 / Open, Classified

RAYSEARCH LABORATORIES AB began a recall of RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs on . FDA classified it as Class II and gives the reason as "A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1482-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
RaySearch Laboratories AB (publ)
Product
RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
FDA root cause
Under Investigation by firm
Action
On February 26, 2024, the firm notified customers of the issue via Urgent Field Safety Notice: Medical Device Correction letters via email. Actions to be taken by the user: 1. Educate planning staff and all users about the possibility of the described use error. 2. If the use error is deemed likely to occur at your clinic, take actions to ensure that proper review of assigned materials is performed before a patient model is used for treatment planning. 3. If you wish to have the Silicon chemical element removed from the material list, contact RaySearch support for assistance. This is especially recommended for clinics that perform ion beam treatments, where the harm caused by the use error could be significant. 4. Inspect your product and identify all installed units with the above software version number(s). 5. Confirm you have read and understood this notice by replying to the notification email. The Silicon chemical element will not be included in the next version of RayStation/RayPlan, scheduled for market release in April 2024 (subject to market clearance in some markets). If requested by the customer, it can be removed from the material list of existing installations. This notice needs to be passed on to all those who need to be aware within your organization. Maintain awareness of this notice as long as any affected version is in use. Thank you for your cooperation, and we apologize for any inconvenience. For regulatory information, please contact [email protected].
Quantity in commerce
154 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hong Kong S.A.R., Hungary, India, Indonesia, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zealand, Norway, Poland, P.R. China, Qatar, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom.
Product codes and lots
Version Number (Build Number/UDI): 8B (8.1.0.47/07350002010129), 8B SP1 (8.1.1.8/07350002010204), 8B SP2 (8.1.2.5/07350002010235), 9A (9.0.0.113/07350002010174), 9A SP1 (9.0.1.142/07350002010488), 9B (9.1.0.933/07350002010266), 9B SP1 (9.2.0.483/07350002010297), 10A (10.0.0.1154/07350002010303), 10A SP1 (10.0.1.52/07350002010365), 10A SP2 (10.0.2.10/07350002010655), 10B (10.1.0.613/07350002010310), 10B SP1 (10.1.1.54/07350002010471), 11A (11.0.0.951/07350002010389), 11A SP1 (11.0.1.29/07350002010433), 11A SP2 (11.0.3.116/07350002010440), 11A SP3 (11.0.4.15/07350002010631), 11B (12.0.0.932/07350002010426), 11B SP1 (12.1.0.1221/07350002010495), 11B SPC1 (12.1.1.41/07350002010587), 11B SP2 (12.0.3.68/07350002010501), 11B SPC2 (12.1.2.91/07350002010617), 11B SP3 (12.0.4.12/07350002010600), 11B SPC3 (12.1.3.162/07350002010662), 11B SPC4 (12.1.4.113/07350002010709), 11B SPC5 (12.1.5.60/07350002010747), 11B SPT1 (12.3.0.119/07350002010570), 12A (13.0.0.1547/07350002010549), 12A SP1 (13.1.0.144/07350002010679), 12A SP2 (13.1.1.89/07350002010730), 2023B (14.0.0.3338/07350002010556), 2024A (15.0.0.430/07350002010723).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1482-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1482-2024Class IIOngoingVoluntary: Firm initiated