Z-1352-2024Class I
Mojo Non Vented Full Face Mask with Headgear, All Sizes
SleepNet Corporation / initiated 2024-03-01 / Open, Classified / root cause: design
SleepNet Corporation began a recall of Mojo Non Vented Full Face Mask with Headgear, All Sizes on . FDA classified it as Class I and gives the reason as "Update to contraindications and warning language due to CPAP masks containing magnets". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1352-2024
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- SleepNet Corporation
- Product
- Mojo Non Vented Full Face Mask with Headgear, All Sizes
- Product code
- BZD Ventilator, Non-Continuous (Respirator)
- Reason for recall
- Update to contraindications and warning language due to CPAP masks containing magnets.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm emailed consignees a notification titled URGENT: MEDICAL DEVICE RECALL, dated 3/1/24. Consignees were instructed to stop using recalled masks if the user or anyone in close physical contact with the mask has an active medical implant or metallic implant that will interact with magnets and contact their mask provider for a replacement mask that does not contain magnets. Consignees that do not have an active medical implant or metallic implant that will interact with magnets do not have to cease use. If a consignee is unsure whether or not to cease use, they should consult with their physician or manufacturer of the implant. Consignees are to continue using masks according to the updated instructions and labeling if patients or people in close proximity to them do not have implanted metallic medical devices or metallic objects in the body. Consignees with any questions are to contact Himabindu Koppusetty via email at [email protected] or by phone at 1-800-742-3646x6611 from Monday through from, 8:00 AM to 4:00 PM.
- Quantity in commerce
- 11,874 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of AUSTRIA, BAHRAIN, BRAZIL, COLOMBIA, COSTA RICA, ECUADOR, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, INDIA, IRAQ, IRELAND, ISRAEL, ISREAL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, MALAYSIA, MEXICO, MOROCCO, NORWAY, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SULTANATE OF OMAN, SWEDEN, SWITZERLAND, THE NETHERLANDS, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VIETNAM, YEMEN.
- Product codes and lots
- All UDI-DI; All Lot Numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1351-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1352-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1353-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1354-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1355-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1356-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1357-2024 | Class I | Ongoing | Voluntary: Firm initiated |