Z-1548-2024Class II

Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to...

OLYMPUS Corporation of the Americas / initiated 2024-03-01 / Open, Classified / root cause: labeling and UDI

Olympus Corporation of the Americas began a recall of Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the... on . FDA classified it as Class II and gives the reason as "Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1548-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract Model: KD-645L
Product code
KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Reason for recall
Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury
Root cause tag
labeling and UDI
FDA root cause
Use error
Action
Olympus Corporation of the Americas issued Medical Device Safety Reminder letter on 3/1/24 via UPS. Letter states reason for recall, health risk and action to take: 1. Follow your facility's procedures for communication and handling of Field Safety Notices. Ensure all personnel, including clinical staff, are informed of the contents of this letter and the Instructions for Use. You may add a copy of this letter with your IFU. 2. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall. b. Enter the recall number: 0443 c. Complete the form as instructed. 3. If you have distributed these devices outside your facility, please provide a copy of this letter to those facilities immediately. Olympus requests you to report any complaints, including any injuries associated with Triangle Tip Electrosurgical knife tip breakage, the Olympus Technical Assistance Center at 1-800-848-9024, option 1. If you require additional information, please do not hesitate to contact me directly at 647-999-3203 or [email protected].
Quantity in commerce
2525 units
Distribution
Nationwide
Product codes and lots
UDI-DI: - 04953170407857 All lots numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1548-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1547-2024Class IIOngoingVoluntary: Firm initiated
Z-1548-2024Class IIOngoingVoluntary: Firm initiated