Z-1373-2024Class II
VITROS Performance Verifier II
Ortho-Clinical Diagnostics, Inc. / initiated 2024-02-27 / Open, Classified
Ortho-Clinical Diagnostics began a recall of VITROS Performance Verifier II on . FDA classified it as Class II and gives the reason as "The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1373-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Performance Verifier II
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.
- FDA root cause
- Under Investigation by firm
- Action
- Consignees were sent an URGENT PRODUCT CORRECTION NOTIFICATION dated 2/27/24 informing them that for product Lot Q1174, diluent Lot M9815 was found in some sales units instead of the expected diluent Lot Q1175, and for product Lot R1176, diluent Lot P9967 was found in some sales units instead of the expected diluent Lot R1177. Consignees were asked to inspect their inventory to see if any mismatched product from affected lots was present and to discard any affected units. The provided Confirmation of Receipt form is to be returned by 3/8/24. Results received using affected devices are to be reviewed for any unexpected QC shift. The recall notification is to be forwarded if any devices were further distributed and the notification maintained in the User Documentation. Customers with any questions can call Global Services at 1-800-421-3311.
- Quantity in commerce
- 11,224 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
- Product codes and lots
- Product Code: 8231474; GTIN: 10758750004577; Lot No. R1176.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1372-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1373-2024 | Class II | Ongoing | Voluntary: Firm initiated |