Z-1399-2024Class II

BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System)...

Becton Dickinson & Co / initiated 2024-02-29 / Open, Classified

Becton Dickinson & Co. began a recall of BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells... on . FDA classified it as Class II and gives the reason as "Esculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However, A17 is checked for...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1399-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Becton Dickinson & Co
Product
BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with carbon sources for yeast identification.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
Esculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However, A17 is checked for fluorescence and if anything is found, the system will abort the panel with the message Fluorescent interference has been detected. Esculin is expected to trigger this fluorescence and test cancellation.
FDA root cause
Under Investigation by firm
Action
BD issued URGENT: Medical Device Product Correction Product Removal letter on 2/29/24 via courier service via tracked letter and/or e-mail. Letter states reason for recall, health risk and action to take: 1. Immediately inspect your inventory for the specific catalog and lot number listed above. Destroy affected product subject to the recall following your institution s process of destruction. 2. Ensure the contents of this notification are read and understood. 3. Share and post this customer letter within your facility network and forward to any customers to whom you may have distributed the product to ensure awareness. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements. BD will issue product replacements following receipt of the completed Customer Response Form. Contact: North American Regional Complaint Center Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: [email protected] Product Advisory, Recall Questions, Product Complaints, Technical Questions
Quantity in commerce
6700 eaches
Distribution
Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.
Product codes and lots
UDI-DI: (10) 3312180 (17) 241030 (30)0025 (01) 30382904483167 Lot Number: 3312180 Exp. Date: 30-OCT-2024

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1399-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1398-2024Class IIOngoingVoluntary: Firm initiated
Z-1399-2024Class IIOngoingVoluntary: Firm initiated