Z-1400-2024Class II

Hill-Rom PRO+ 36" MRS Surface, REF P7924A03, The pro+ mattress is intended for patient support and for the...

Baxter Healthcare Corporation / initiated 2024-03-05 / Open, Classified

Baxter Healthcare Corporation began a recall of Hill-Rom PRO+ 36" MRS Surface, REF P7924A03, The pro+ mattress is intended for patient support and for the prevention and/or treatment of pressure injuries on . FDA classified it as Class II and gives the reason as "Inconsistencies were identified with service records associated with corrections performed for a previous Medical Device Correction issued by Baxter (ref number: FA-2022-026). Your devices have been identified as...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1400-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
Hill-Rom PRO+ 36" MRS Surface, REF P7924A03, The pro+ mattress is intended for patient support and for the prevention and/or treatment of pressure injuries.
Product code
IOQ Bed, Flotation Therapy, Powered
Reason for recall
Inconsistencies were identified with service records associated with corrections performed for a previous Medical Device Correction issued by Baxter (ref number: FA-2022-026). Your devices have been identified as potentially impacted by these service inconsistencies and therefore, Baxter will need to reassess the impacted mattresses to confirm they have been corrected as identified by the service order, and to confirm the correct serial numbers have been identified.
FDA root cause
Under Investigation by firm
Action
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 03/05/2024 via USPS First Class Mail. The notice explained the issue, hazard involved, and requested the consignee inspect the device to check for any indication of error present. If an error is detected, contact Baxter Technical Support by phone at 800-445-3720, available Monday through Thursday, between 8:00 am and 6:30 pm Eastern Time, and Friday, between 8:00 am and 6:00 pm Eastern Time, or by email at [email protected] to have the pro+ mattress corrected. A Baxter representative will contact your facility to arrange for an inspection of your device and correction, if required.
Quantity in commerce
232 units
Distribution
US Nationwide distribution in the state of Colorado.
Product codes and lots
UDI/DI 00887761977884, Serial Numbers: W330BP6363, W333BP6439, W334BP6477, W334BP6535, W337BP6631, W341BP6675, X047BP7353, X049BP7431, W330BP6365, W333BP6440, W334BP6478, W334BP6536, W337BP6632, W341BP6676, X047BP7354, X049BP7432, W330BP6366, W333BP6442, W334BP6480, W334BP6537, W337BP6633, W341BP6678, X047BP7355, X049BP7433, W330BP6367, W333BP6445, W334BP6481, W337BP6565, W337BP6634, W341BP6679, X047BP7356, X063BP7505, W330BP6368, W333BP6446, W334BP6485, W337BP6566, W337BP6635, W341BP6681, X047BP7357, X063BP7506, W330BP6382, W333BP6447, W334BP6486, W337BP6567, W337BP6636, W341BP6683, X047BP7358, X063BP7507, W330BP6383, W333BP6448, W334BP6487, W337BP6568, W337BP6637, W341BP6684, X047BP7359, X063BP7508, W330BP6385, W333BP6450, W334BP6489, W337BP6569, W337BP6638, W341BP6687, X047BP7360, X063BP7509, W330BP6387, W333BP6451, W334BP6492, W337BP6570, W337BP6641, X046BP7330, X047BP7361, X063BP7510, W330BP6397, W333BP6452, W334BP6493, W337BP6571, W337BP6642, X046BP7331, X047BP7362, X063BP7511, W333BP6399, W333BP6453, W334BP6496, W337BP6572, W337BP6643, X046BP7332, X047BP7363, X063BP7512, W333BP6400, W333BP6455, W334BP6500, W337BP6573, W337BP6644, X046BP7333, X047BP7364, X063BP7513, W333BP6401, W333BP6456, W334BP6501, W337BP6574, W337BP6645, X046BP7334, X047BP7365, X063BP7514, W333BP6404, W333BP6457, W334BP6502, W337BP6575, W341BP6646, X046BP7335, X047BP7366, X066BP7516, W333BP6405, W333BP6458, W334BP6504, W337BP6577, W341BP6647, X046BP7336, X047BP7367, X066BP7517, W333BP6406, W333BP6459, W334BP6505, W337BP6578, W341BP6648, X046BP7337, X047BP7368, X066BP7518, W333BP6415, W333BP6460, W334BP6511, W337BP6579, W341BP6655, X046BP7338, X047BP7369, X066BP7519, W333BP6418, W333BP6461, W334BP6513, W337BP6583, W341BP6658, X046BP7339, X047BP7370, X066BP7520, W333BP6420, W333BP6463, W334BP6514, W337BP6614, W341BP6659, X046BP7340, X047BP7371, X066BP7521, W333BP6421, W333BP6465, W334BP6525, W337BP6615, W341BP6660, X047BP7342, X047BP7372, X066BP7522, W333BP6422, W333BP6466, W334BP6526, W337BP6621, W341BP6661, X047BP7343, X047BP7373, X066BP7523, W333BP6423, W333BP6467, W334BP6527, W337BP6622, W341BP6663, X047BP7344, X047BP7374, X066BP7524, W333BP6424, W334BP6468, W334BP6528, W337BP6623, W341BP6666, X047BP7345, X047BP7375, X066BP7525, W333BP6425, W334BP6469, W334BP6529, W337BP6624, W341BP6667, X047BP7346, X047BP7376, X066BP7526, W333BP6426, W334BP6470, W334BP6530, W337BP6625, W341BP6668, X047BP7347, X047BP7378, X066BP7527, W333BP6427, W334BP6471, W334BP6531, W337BP6626, W341BP6669, X047BP7348, X047BP7379, X066BP7531, W333BP6431, W334BP6473, W334BP6532, W337BP6628, W341BP6670, X047BP7350, X047BP7381, X066BP7532, W333BP6432, W334BP6475, W334BP6533, W337BP6629, W341BP6671, X047BP7351, X049BP7429, X066BP7533, W333BP6435, W334BP6476, W334BP6534, W337BP6630, W341BP6674, X047BP7352, X049BP7430, X066BP7540

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1400-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1400-2024Class IIOngoingVoluntary: Firm initiated