Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1329-2024Class II

HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the...

Baxter Healthcare Corporation / initiated 2024-03-11 / Open, Classified

Baxter Healthcare Corporation began a recall of HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery on . FDA classified it as Class II and gives the reason as "Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1329-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery
Product code
HST Apparatus, Traction, Non-Powered
Reason for recall
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.
FDA root cause
Under Investigation by firm
Action
Baxter issued an Urgent Medical Device Recall notice to its sole consignee on 03/11/2024 via email. The notice explained the issue and potential hazard and requested the following: 1. Immediately locate, isolate, and cease all use of the affected serial numbers of the product. The product code and serial number can be found on labelling on the bottom of the boot. 2. Arrange for the return of your impacted product by calling the Baxter Order Management Team at 800-433-5774, option 2 then 3, between the hours of 8:00 am and 5:00 pm Eastern Time, Monday through Friday, or by emailing Baxter at [email protected]. 3. Acknowledge the receipt of this notification by completing a reply form on the customer portal.
Quantity in commerce
80 units
Distribution
US Nationwide distribution in the state of Florida.
Product codes and lots
123510000293, 123510000294, 123510000295, 123510000296, 123510000298, 123510000299, 123510000300, 123510000764, 123510000765, 123510000766, 123510000767, 123510000768, 123510000769, 123510000770, 123510000771, 123510000772, 123510000773, 123510000800, 123510000801, 123510000802, 123510000803, 123510000804, 123510000805, 123510000806, 123510000807, 123510000808, 123510000809, 123510000297, 123350001033, 123350001034, 123350001035, 123350001036, 123350001040, 123350001041, 123350001042, 123360000294, 123360000295, 123360000297, 123360000298, 123360000299, 123360000300, 123360000301, 123360000302, 123360000303, 123360000304, 123360000733, 123360000734, 123360000735, 123360000736, 123360000737, 123360000738, 123360000739, 123360000740, 123360001972, 123360001973, 123360001974, 123360001975, 123360001976, 123360001977, 123360001978, 123360001979, 123360001980, 123360001981, 123350000968, 123350000969, 123350000970, 123350000971, 123350000972, 123350000973, 123350000974, 123350000975, 123350000987, 123350000988, 123350000989, 123350000990, 123350000991, 123350000992, 123350000993, 123350000994, 123350001032

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1329-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1329-2024Class IIOngoingVoluntary: Firm initiated