Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1601-2024Class II

ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running...

Siemens Medical Solutions USA, Inc. / initiated 2024-02-23 / Open, Classified / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A on . FDA classified it as Class II and gives the reason as "On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1601-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
Root cause tag
software
FDA root cause
Software design
Action
On 2/23/24, correction notices were sent to customers informing them of the following: This issue is specific to and limited to the use of the Cardiac DICOM SR feature: - When a user uses and configures the Cardiac DICOM SR feature to display either the MINIMUM (Min) or the MAXIMUM (Max) measured value, AND - Makes multiple measurements of the cardiac region during an examination, AND - Exports the results into the Cardiac DICOM SR feature, THEN - The Cardiac DICOM SR viewer will NOT display the MINIMUM (Min) or MAXIMUM (Max) value and will instead display the LAST measured value. In that case, the displayed measurement in the Cardiac DICOM SR viewer may NOT be the Minimum (Min) or MAXIMUM (Max) value measured during the examination. To avoid potential misdiagnosis by an over- or underrepresented measured value, it is recommended to refer to the measurement values displayed on the Ultrasound System Final Report when the configuration is to display the Min or Max value. Firm recommends a review of previous cardiac ultrasound examination results where the clinical assessment was reviewed using a DICOM SR viewer. Your Customer Service Engineer from will contact you to schedule a facility visit to update the system or inform you of a remote update when the software update is available. The software update is currently under development and estimated to be available by spring/summer of 2024. Please ensure that all users of the affected devices within your organization, and others who may need to be informed, receive the safety relevant information provided with this notice and take the actions specified herein. If you have further questions, please contact the firm's Ultrasound Service Customer Care Center at 1-800-888-7436. Not Affected: -ACUSON Juniper Ultrasound System, product version 2.5, software version VB30D. - ACUSON Redwood system, product version 2.0.
Quantity in commerce
67
Distribution
Worldwide distribution - US Nationwide including in the states of AL, AR, CA, CO , CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA , MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WA, WI, WY and the countries of BG, BW, BA, TR, AR, CL, CN, FR, SK, CD, BH, PH, AM, CR, AF, BE, AE, AZ, AO, KR, RO, HU, HR, DK, AU, DZ, CY, PL, BO, AL, NL, CI, BR, CZ, DE, AT, ES, CO, BD, CH, IT, CA, CM, DO, EC, EG, ER, ET, FI, FO, GB, GE, GR, GT, GY, HK, HN, ID, IE, IL, IN, IQ, IR, JO, JP, KE, KW, KZ, LB, LT, LU, LV, LY, MA, MD, ME, MK, MM, MN, MU, MW, MX, MY, NG, NO, NP, NZ, OM, PA, PE, PG, PK, PT, PY, QA, RS, RU, SA, SE, SG, SI, SM, SN, SV, SZ, TH, TN, TT, TW, TZ, UA, UG, UY, UZ, VN, XK, YE, ZA, ZM, ZW.
Product codes and lots
REF/UDI-DI: 11711750/04056869995809, and 11711779/04056869995830. Serial Numbers: 700127 700125 700109 700123 700124 700126 700131 700077 700081 700070 700071 700073 700075 700076 700072 700074 700096 700097 700098 700100 700103 700082 700030 700036 700037 700028 700029 700031 700033 700022 700023 700024 700025 700035 700121 700111 700132 700065 700058 700026 700019 700020 900581 900582 900583 900585 700034 700032 700041 700040 700039 700052 700053 700057 700038 700063 700062 700060 700064 700068 700059 700084 700083 700107 700114 700115 700135.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1601-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1599-2024Class IIOngoingVoluntary: Firm initiated
Z-1600-2024Class IIOngoingVoluntary: Firm initiated
Z-1601-2024Class IIOngoingVoluntary: Firm initiated