RENASYS EDGE 300ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pump...
Smith & Nephew Medical, Ltd. began a recall of RENASYS EDGE 300ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pump (NPWT), as it may promote wound healing. Product... on . FDA classified it as Class II and gives the reason as "Black particulate matter from the carbon filter component of the canisters entering the NPWT pump and potentially causing damage to the pump". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1596-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Smith & Nephew Medical Ltd
- Product
- RENASYS EDGE 300ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pump (NPWT), as it may promote wound healing. Product Number: 66803139
- Product code
- OMP Negative Pressure Wound Therapy Powered Suction Pump
- Reason for recall
- Black particulate matter from the carbon filter component of the canisters entering the NPWT pump and potentially causing damage to the pump
- FDA root cause
- Under Investigation by firm
- Action
- Smith & Nephew Medical Ltd. issued Urgent Medical Device Recall Notice R-2024-02 on March 11, 2024 via overnight mail. Letter states reason for recall, health risk and action to take: Based on the regulatory classification and degree of hazard, this field action extends to the retail level. Therefore, it is not required to communicate this notice to the patient at the consumer level. If your customer contacts you regarding a pump failure: 1. Contact S+N Clinical Hotline (800-876-1261) with any questions. 2. Please ensure you provide a replacement as soon as possible. Replacement Stock. Smith+Nephew is supplying replacement stock quantities in parallel with this notice. Please contact [email protected] if you have any questions. If you have any questions or concerns regarding this recall please contact [email protected].
- Quantity in commerce
- 6895 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Chile.
- Product codes and lots
- GTIN: (01)05000223501044 Lot Numbers: 1016448, 1016478, 1016537, 1016773, 1017147, 1017417, 1017418, 1017912, 1016044
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1596-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1597-2024 | Class II | Ongoing | Voluntary: Firm initiated |