Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098
Intuitive Surgical, Inc. began a recall of Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098 on . FDA classified it as Class II and gives the reason as "Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1518-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 3/13/24, Field Safety Notices were mailed and emailed to customers who were asked to do the following: 1) To prevent a pinch injury, please use two people when attempting to move the EGM in an unpowered state. 2) Inform necessary hospital personnel about this issue and affected personnel when the requested actions have been completed. 3) Complete and return the response form via email to [email protected] A firm representative will schedule a site visit to perform the inspection of the Link 2 EGM belts of the affected population of SP systems. If you need further information or support concerning this Medical Device Notification, please contact your Clinical Sales Representative or contact Firm Customer Service: - North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: [email protected]. - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or [email protected] - South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) or [email protected] - Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) or [email protected]
- Quantity in commerce
- 53
- Distribution
- US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV, MS, TX, KS. OUS: China, France, Japan, South Korea
- Product codes and lots
- Part/UDI-DI: 380601/00886874114605. SP1098 Serial Numbers: SP0202, SP0253, SP0254, SP0219, SP0228, SP0231, SP0238, SP0273, SP0208, SP0226, SP0240, SP0247, SP0249, SP0256, SP0257, SP0268, SP0270, SP0272, SP0221, SP0223, SP0225, SP0229, SP0230, SP0232, SP0233, SP0235, SP0236, SP0237, SP0239, SP0241, SP0242, SP0243, SP0244, SP0245, SP0248, SP0250, SP0251, SP0252, SP0255, SP0258, SP0259, SP0260, SP0261, SP0262, SP0263, SP0264, SP0265, SP0266, SP0269, SP0274, SP0278, SP0280, SP0282
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1518-2024 | Class II | Ongoing | Voluntary: Firm initiated |