Z-1549-2024Class II

Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium...

Siemens Healthcare Diagnostics Inc. / initiated 2024-03-13 / Open, Classified / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using the Atellica CI... on . FDA classified it as Class II and gives the reason as "Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1549-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparin and sodium heparin) using the Atellica CI Analyzer and Atellica CH Analyzer Siemens Material Number (SMN): 11537211
Product code
JIY Photometric Method, Iron (Non-Heme)
Reason for recall
Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts calibrator, quality control (QC), and patient results.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens issued Urgent Medical Device Correction (UMDC) ACHC24-03.A.US to US customers via FedEx beginning on 03/13/2024. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 03/13/2024. New customers who purchase the Atellica CH Iron3 assay on the Atellica CH and Atellica CI analyzers will receive the UMDC/UFSN, until the product issue is resolved. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Refer to short-term mitigation steps in Appendix B while Siemens works towards transitioning customers back to the Iron_2 assay. {The UMDC/UFSN provides instructions to Atellica Solutions customers with multiple analyzers to segregate Iron3 and the impacted assays (Chol_2, LDLC, and Trig_2) onto separate systems. Atellica Solutions customers with a single Atellica analyzer instructed to stop use of Iron3 and identify alternative methods for Iron testing]. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution Siemens is actively working to ramp up the supply volume to support customer conversion to the Iron_2 assay. We anticipate being able to support an end to the mitigation within a six month timeframe. A follow up communication will be provided when Customer Actions are no longer required. These mitigations will remain in effect until the solution is provided
Quantity in commerce
8721 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Curacao, St¿Eus, Czech Republic, Denmark, Estonia, Finland, France, Germany, ¿ Greece, Hungary, India, Ireland, Israel, Italy, Latvia, Lithuania, Mexico, Netherlands, ¿¿¿¿ New Zealand, Norway, Paraguay, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., United¿Kingdom.¿
Product codes and lots
UDI-DI: 00630414610849 All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1549-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1549-2024Class IIOngoingVoluntary: Firm initiated