Z-1520-2024Class I

AW-S030 NEO SONARMED SENSOR 3.0MM. Used in conjunction with SonarMed Monitor for airway monitoring

Sonarmed, Inc. / initiated 2024-03-21 / Open, Classified / root cause: manufacturing process

SonarMed Inc began a recall of AW-S030 NEO SONARMED SENSOR 3.0MM. Used in conjunction with SonarMed Monitor for airway monitoring on . FDA classified it as Class I and gives the reason as "The failure to detect the partial obstruction in a 2.5mm sensor". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1520-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Sonarmed, Inc.
Product
AW-S030 NEO SONARMED SENSOR 3.0MM. Used in conjunction with SonarMed Monitor for airway monitoring.
Product code
OQU Airway Monitoring System
Reason for recall
The failure to detect the partial obstruction in a 2.5mm sensor.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
An Urgent Medical Device Recall notification dated 3/25/2024 was mailed to consignees via UPS notifying them of this SonarMed Sensor recall. Consignees were asked to discontinue use of affected devices, quarantine any affected devices, and return all quarantined devices to Medtronic by obtaining a Returned Goods Authorization (RGA) from [email protected]. If product was purchased through a distributor the distributor can arrange for return. The provided notification is to be provided to those within consignee organizations that need to be made aware and if product was further distributed. Consignees with questions are to contact customer service at 800-962-9888, Option 2.
Quantity in commerce
690 units
Distribution
Domestic: AK, AZ, CA, FL, IL, KY, MD, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, & WA.
Product codes and lots
Product Number/CFN: AW-2030; UDI-DI: 10851334007190; Serial Numbers: A20221012, A20221017, A20221024, A20221103, A20221115, A20221116, A20221117, A20221118, A20221130, A20221206, A20221222, A20221228, A20221229, A20230104, A20230111, A20230116, A20230117, A20230118, A20230208, A20230209, A20230313, A20230314, A20230315, A20230316, A20230317, A20230405.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1520-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1519-2024Class IOngoingVoluntary: Firm initiated
Z-1520-2024Class IOngoingVoluntary: Firm initiated
Z-1521-2024Class IOngoingVoluntary: Firm initiated