Z-1535-2024Class I
AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring
Sonarmed, Inc. / initiated 2024-03-25 / Open, Classified / root cause: design
SonarMed Inc began a recall of AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring on . FDA classified it as Class I and gives the reason as "Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1535-2024
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Sonarmed, Inc.
- Product
- AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring.
- Product code
- OQU Airway Monitoring System
- Reason for recall
- Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An Urgent Medical Device Recall notification dated 3/25/2024 was mailed to consignees via UPD notifying them of this SonarMed Sensor and Monitor recall. Consignees were asked to discontinue use of affected devices, quarantine any affected devices, and return all quarantined devices to Medtronic by obtaining a Returned Goods Authorization (RGA) from [email protected]. If product was purchased through a distributor the distributor can arrange for return. The provided notification is to be provided to those within consignee organizations that need to be made aware and if product was further distributed. Consignees with questions are to contact customer service at 800-962-9888, Option 2.
- Quantity in commerce
- 145 units
- Distribution
- _AK, AZ, CA, FL, IL, KY, MD, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, & WA_
- Product codes and lots
- Model No. AW-M0001; GTIN: 00851334007001; Serial No. AW0572, AW0573, AW0574, AW0577, AW0578, AW0579, AW0580, AW0581, AW0584, AW0585, AW0586, AW0588, AW0591, AW0593, AW0594, AW0600, AW0606, AW0607, AW0615, AW0616, AW0624, AW0626, AW0627, AW0648, AW0650, AW0651, AW0658, AW0664, AW0672, AW0674, AW0675, AW0678, AW0680, AW0681, AW0684, AW0685, AW0701, AW0702, AW0703, AW0704, AW0705, AW0706, AW0707, AW0708, AW0709, AW0710, AW0711, AW0712, AW0718, AW0719, AW0720, AW0721, AW0722, AW0723, AW0724, AW0725, AW0726, AW0727, AW0728, AW0729, AW0730, AW0731, AW0732, AW0733, AW0734, AW0735, AW0736, AW0737, AW0738, AW0739, AW0740, AW0741, AW0742, AW0743, AW0744, AW0745, AW0746, AW0747, AW0748, AW0749, AW0763, AW0764, AW0765, AW0766, AW0767, AW0768, AW0769, AW0770, AW0771, AW0772, AW0773, AW0774, AW0775, AW0776, AW0777, AW0778, AW0779, AW0780, AW0781, AW0782, AW0783, AW0784, AW0785, AW0786, AW0787, AW0788, AW0789, AW0790, AW0791, AW0792, AW0793, AW0794, AW0795, AW0796, AW0797, AW0798, AW0799, AW0800, AW0804, AW0805, AW0814, AW0817, AW0818, AW0819, AW0820, AW0821, AW0825, AW0826, AW0827, AW0828, AW0830, AW0831, AW0832, AW0833, AW0834, AW0835, AW0836, AW0837, AW0838, AW0839, AW0840, AW0841, AW0842, AW0854, AW0855.