5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500...
Philips Ultrasound, Inc. began a recall of 5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300 W/795146, 5300... on . FDA classified it as Class II and gives the reason as "Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1757-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- 5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300 W/795146, 5300 P/795147, 5300 G/795148, used with Deluxe and Premium Carts that come with multiport adapters.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer connected 6) Non-TEE imaging continues, Then though TEE is unselected, it may have power, no temperature monitoring/control, and may result in temperature above limit and patient burns
- FDA root cause
- Component change control
- Action
- On 3/23/24, correction notices were distributed to customers who were advised of the following two ways to avoid the condition: Option 1: To prevent the issue from occurring, when using the TEE transducers, connect directly to the Compact 5000 system using the port on the side of the system as shown in Figure 3 below. The issue is associated with the MPA, therefore by using the port on the Compact 5000 directly, the issue is avoided. Option 2: Whenever adding a transducer to, or removing a transducer from the MPA, disconnect and reconnect the MPA connector on the Compact 5000 system connector port, followed by a transducer selection on the touch screen. By performing this step when adding or removing a transducer, the monitoring and control systems are re-initialized. Circulate this URGENT Medical Device Correction Letter to all users of this device within your organization and forward to any organization where potentially affected devices have been transferred so that they are aware of the issue. In addition, customers were asked to complete and return the response form [email protected] A firm representative will contact you to schedule a time for a Field Service Engineer to visit and implement the solution to resolve the issue For further information or support concerning this issue, please contact your local firm representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).
- Quantity in commerce
- 363
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, CO, CT, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, NC, NH, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA and the countries of Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Lithuania, Netherlands, New Zealand, Oman, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates.
- Product codes and lots
- All Compact 5000 series systems manufactured on or before 20-Dec-2023 and purchased with Deluxe and Premium Cart options. REF/UDI-DI/Serial Numbers: 505500 CV/795141/00884838116566/USN23P0684, USN23P0623, USN23P0624, USN23P0685, USN23P0720, USN23P0615, USN23P0627, USN23P0611, USN23P0683, USN23P0682, USN23P0625, USN23P0173, USN23P0164, US823P0231, USN23P0613, USN23P0172, USN23P0721, USN23P0410, USN23P0612, USN23P0614, USN23P0171, US623P0141, US823P0233, US423P0844, US523P1060, US523P1061, US523P1750, US523P2536, US523P2537, US523P2672, US523P2673, US423P0842, US623P0140, US423P0759, US623P0142, US623P0230, US623P0485, US623P0486, US623P0740, US623P0790, US623P0791, US623P0792, US623P0866, US623P0867, US423P0843, US623P0966, US423P0672, US423P0673, US423P0674, US423P0712, US423P0757, US823P0132, US823P0232, US823P0318, US823P0319, US823P0320, US823P0321, US823P0322, US823P0507, US823P0508, US823P1857, US923P1161, US823P0129, US723P1919, US623P1072, US623P1403, US623P1483, US623P1606, US623P1607, US623P1608, US623P1609, US823P0131, US823P0029, US823P0128, US823P0127, US823P0033, US823P0032, US823P0030, US823P0028, US723P2256, US723P2154, US723P2153, US723P2152, US823P0130, US823P0031, USO23P2173, USO23P2065, US723P1242, US723P1243, US723P1342, US723P1343, US723P1241, US723P1240, USN23P0333, USN23P0332, US623P1450, US723P1766, USO23P0157, USO23P0156, USO23P0059, USO23P0058, US823P1743, US823P1744, US823P1745, US723P1582, US723P1872, US723P1583, USO23P2274, USO23P2064, USO23P2171, US623P1739, US723P1239, US723P1111, USN23P0330, USN23P0331, US923P0205, USN23P0051, US823P1767, US923P0207, US823P1770, US823P1769, US823P1251, US823P1229, US723P0393, US923P0421, USO23P2069; 5500 P/795143/00884838116573/USN23P0048, USN23P0050, USO23P2276, USO23P2275, US523P2372, US523P2538, US523P2406, US523P2405, US523P2374, US523P2373, US523P1837, US523P1838, US623P0741, US523P1839, US823P0744, US823P0745, US723P2257, US823P1030, US823P0877, US823P0548, US823P0547, US823P0546, US823P0511, US823P0509, US823P0317, US723P2258, US823P0510, US623P1486, US623P1485, US623P1741, US623P0969, US623P0968, US723P0809, US823P2055, US823P2054, US723P1584, US723P1472, US723P1471, USN23P0334, US823P2053, US723P1917, US923P0381, US823P1254, US723P0182, US723P0063, US623P2118, USO23P0171, US723P1918, US823P1250, USO23P2174, US723P0171, US723P0172, US623P2117, US723P0061, US723P0062, US723P0170, USN23P0753, US723P1809. US922P0888. US723P0596, US623P1738, US723P1112, US723P1113, US723P1767; 5500 W/795144/00884838116580/USN23P0628, USN23P0411, USN23P0174, USN23P0166, US523P2404, US623P0603, US623P0601, US423P0994, US423P0993, US423P0995, US623P0602, US623P2004, US623P1075, US723P0174, US723P0173, US623P2109, US623P2005, US922P0887, US623P1740; 5500 G/795145/00884838116597/USN23P0165, USN23P0053, USO23P1901, USO23P2068, USN23P0686, USN23P0414, USN23P0413, USN23P0412, USN23P0335, US523P2674, US523P2800, US523P2798, US523P2799, US623P0318, US623P0395, US623P0394, US623P0393, US623P0319, US623P0229, US623P0228, US623P0019, US623P0018, US623P0017, US623P0320, US523P2152, US523P2151, US523P2014, US523P2013, US523P2012, US523P1749, US523P1440, US523P1439, US823P0984, US823P0742, US823P0743, US823P0730, US823P0879, US823P1084, US823P0550, US823P0512, US823P0732, US723P2261, US723P2260, US723P2259, US623P1281, US623P1610, US623P1404, US623P1280, US623P1713, US823P0549, US823P1853, US823P1855, US823P1856, US823P1854, US723P1473, US723P1474, US723P1344, US723P1346, US723P1345, US723P1475, US823P1676, US823P0505, US723P2155, US723P2156, US823P1652, USN23P0626, US823P1678, US723P2157, USN23P0170, US723P1765, US623P2119, US823P1230, US723P0183, USO23P0172, USO23P0173, USO23P0174, US823P1677, US723P1763, US723P0064, US823P1747, US823P1746, US823P0741, USO23P2175, US723P1585, US923P0204, US723P0390, US723P0185, US723P0184, US723P0391, US723P0259, US723P0186, US723P0392, US723P0260, US723P0258, US723P0257, US723P0256, US923P0419, USN23P0049, US423P0714, US523P0946, US223P0651, US723P0599, US723P0598, US723P0597, US823P0880, US223P0650, US623P1863, US623P1864, USO23P2279, US723P1114, US723P1115, US723P1914, US723P1915, US723P1916, US623P2000, USO23P1805, US823P1768, USO23P2277, US823P2052, US723P0743, US723P0744, US823P2051, US723P0993, US723P0992, US923P0206, US823P1232, US923P0420, US723P1764, US823P1231, US923P0208; 5300 W/795146/00884838116603/US922P0893, USN23P0616, USN23P0177; 5300 P/795147/00884838116610/USN23P0724, US922P0892, USN23P0415, US723P0850, US823P0836, US723P0810, US623P0908, US623P2124, US623P2123; 5300 G/795148/00884838116627/US922P0891, US823P0230, US823P0228, US623P0484, US823P0229, USO23P2070, USO23P2176, US723P1586, USN23P0175, USN23P0176, US723P0600, USN23P0052
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1757-2024 | Class II | Ongoing | Voluntary: Firm initiated |