Z-1696-2024Class II

OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression...

Advanced Instruments, LLC / initiated 2024-03-11 / Open, Classified / root cause: manufacturing process

Advanced Instruments, LLC began a recall of OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and... on . FDA classified it as Class II and gives the reason as "System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1696-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Advanced Instruments, LLC
Product
OsmoPRO MAX Automated Osmometer- IVD Osmometer for Clinical Use. Uses the method of freezing point depression to measure osmolality of serum, plasma, and urine. Model/Catalog Number: OsmoPRO MAX Software Version: 1.5.0.0
Product code
JJM Osmometer For Clinical Use
Reason for recall
System error messages may delay the ability to test patient samples and the need for frequent calibration due to results outside the expected range when testing controls.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Advanced Instruments issued Urgent Medical Device Recall Letter on 3/11/24 via email. Letter states reason for recall, health risk and action to take: 1. Discontinue use of your device. 2. Advanced Instruments will contact you regarding the return and exchange of your Device. Several solutions are available as needed to ensure continuity of testing. 3. For those users whose lab protocol requires the review of previous data, Advanced Instruments is available to review the data. 4. Complete and return the attached acknowledgement form. Advanced Instruments is implementing permanent corrective actions to address the two issues described and as part of the process above you will be provided with information on the next steps to implement a solution for impacted devices. If you need any further information or support concerning this issue, please contact [email protected]. To better assist you, please have your instrument serial number ready. 1-800-225-4034 (Toll Free US & Canada) Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.
Quantity in commerce
58 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, France, Germany, Iceland, Netherlands, South Korea, Spain, Switzerland, Taiwan.
Product codes and lots
UDI-DI: 00816068021150; Serial Numbers: 23030208A, 23050414A, 23050417A, 23010059A, 23040283A, 23010014A, 23020166A, 23040284A, 23020159A, 23050410A, 23030236A, 23020165A, 23020156A, 23020161A, 23030239A, 23030210A, 23030233A, 23030232A, 23020158A, 23030206A, 22111360A, 23020155A, 23030209A, 23040287A, 22111362A, 23020160A, 23040288A, 23010052A, 22111354A, 23010055A, 22111350A, 22111365A, 23010053A, 22111363A, 23030237A, 23010062A, 23010015A, 23030202A, 23030207A, 23040290A, 23030204A, 23020157A, 22111367A, 23020158A, 23030203A, 23030240A, 22111358A, 22111359A, 23040286A, 22111355A, 23030205A, 23020164A, 22111368A, 23030238A, 23010060A, 23010061A, 23020163A, 23010011A22111364A. All serial numbers that end in A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1696-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1696-2024Class IIOngoingVoluntary: Firm initiated