Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT...
Philips North America Llc began a recall of Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterruptable Power Supply... on . FDA classified it as Class II and gives the reason as "Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1690-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterruptable Power Supply Model: SCL500RM1U 2. APC (Schneider Electric IT Corporation) 230V Uninterruptable Power Supply Model: SCL500RMI1U
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the delay in the detection of a change in a patient condition
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Philips issued Urgent Medical Device Correction Letter on 3/22/24. Letter states reason for recall, health risk and action to take: If your UPS fails : 1) Promptly take the UPS offline and plug any devices that had been attached to it into a different appropriate power source. Note: It is recommended that UPS devices intended for operation in network and server rooms be kept in a temperature and humidity controlled environment, ensuring adequate airflow around the UPS to prevent UPS device failure. This notice should be passed on to all those who need to be aware within your organization or to any organization where affected devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will replace the malfunctioning UPS. If you need any further information, please contact your local Philips representative, or call 1-800- 722- 9377. Philips Reference# 2024-CC-HPM-010
- Quantity in commerce
- 12,936 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, Estonia, Fiji, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Oman, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, United Kingdom, Utd. Ara Emir.
- Product codes and lots
- 1. SCL500RM1U (only serial number date codes prior to 2237*) 2. SCL500RMI1U (only serial number date codes prior to 2241*) Example unit serial number is 5S"2104"08943, meaning it was manufactured during week 04 of the year 2021. The serial number format is 5SYYWW123456.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1690-2024 | Class II | Ongoing | Voluntary: Firm initiated |