Z-1756-2024Class II
The Oncology Essentials Package of the Spectral CT imaging system
Philips North America LLC / initiated 2024-03-29 / Open, Classified
Philips North America Llc began a recall of The Oncology Essentials Package of the Spectral CT imaging system on . FDA classified it as Class II and gives the reason as "Potential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1756-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- The Oncology Essentials Package of the Spectral CT imaging system.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Potential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment.
- FDA root cause
- Under Investigation by firm
- Action
- An URGENT Medical Device Correction Letter, dated 3/27/24, was mailed to consignees to notify them of this recall. Consignees are instructed to identify if a third-party laser was installed on their system and if the laser line is reflected, duplicated, or thickened. If a third-party laser is installed, before treatment of a patient, reposition or re-image as necessary to verify proper placement of the localization laser marks. Systems may continue to be used in accordance with the intended use and provided use recommendations. Consignees are to complete and return the provided response form by email to [email protected]. The provided recall notification is to be disseminated to all users of affected systems. Philips will contact consignees to schedule a time for a Field Service Engineer to visit and install hardware to resolve the issue. Questions are to be directed to the Customer Care Solutions Center at 1-800-722-9377 from 8:00 AM to 8:00 PM, Monday through Friday.
- Quantity in commerce
- 12 units
- Distribution
- Worldwide distribution - US Nationwide in the states of AZ, HI, MN, VT, WV and the countries of China, Macao, New Zealand, Portugal, Switzerland.
- Product codes and lots
- Model No. 728333; UDI: (01)00884838101111(21); Serial No. 10069, 10084, 10169, 10137, 10163, 10196, 10181, 10199, 10210, 10206, 10202, 10139.