Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug...
Randox Laboratories Ltd. began a recall of Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine... on . FDA classified it as Class III and gives the reason as "Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1786-2024
- Classification
- Class III
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Randox Laboratories, Limited
- Product
- Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine Catalog Number: EV4115
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Randox Laboratories Ltd UK notified the US Distributor Randox Laboratories via email on 4/05/ 24. End-Users notified by email on 4/9/24. Letter (REC736) states reason for recall, health risk and action to take: " Export support files for the last 10 samples run and share with Randox using the link to be provided " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to [email protected] within five working days. " Affected chip heater assembly replacement to be arranged by Randox Service Engineer Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation
- Quantity in commerce
- 40 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- GTIN: 05055273214970 Serial Numbers: 052-23-0424 052-23-0425 052-23-0434 052-23-0437 052-23-0618 052-23-0611 052-23-0516 052-23-0522 052-23-0490 052-23-0491 052-23-0439 052-23-0440 052-23-0441 052-23-0445 052-23-0449 052-23-0450 052-23-0451 052-23-0452 052-23-0438 052-22-0276 052-22-0277 052-22-0279 052-23-0416 052-22-0283 052-22-0271 052-21-0260 052-21-0255 052-21-0249 052-23-0418 052-23-0419 052-23-0415 052-23-0417 052-22-0268 052-23-0420 052-22-0284 052-23-0448 052-23-0444 052-23-0443 052-23-0442 052-23-0609
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1786-2024 | Class III | Ongoing | Voluntary: Firm initiated |